Reproducibility of Laser Doppler Flowmetry in gingival microcirculation. A study on six different protocols.

作者信息

Ajan Aida, Roberg Karin, Fredriksson Ingemar, Abtahi Jahan

机构信息

Department of Oral and Maxillofacial Surgery, Linköping University Hospital, Linköping, Sweden; Department of Biomedical and Clinical Sciences (BKV), Linköping University, Linköping, Sweden.

Department of Biomedical and Clinical Sciences (BKV), Linköping University, Linköping, Sweden; Department of Otorhinolaryngology in Linköping, Region Östergötland, Linköping, Sweden.

出版信息

Microvasc Res. 2024 May;153:104666. doi: 10.1016/j.mvr.2024.104666. Epub 2024 Jan 30.

Abstract

OBJECTIVES

Laser Doppler Flowmetry (LDF) is a non-invasive technique for the assessment of tissue blood flow, but increased reproducibility would facilitate longitudinal studies. The aim of the study was to assess the interday reproducibility of Laser Doppler Flowmetry (LDF) at rest, at elevated local temperatures, and with the use of the vasodilator Methyl Nicotinate (MN) in six interconnected protocols for the measurement of the blood supply to the microvascular bed of the gingiva.

METHODS

Ten healthy volunteers were included. Interweek LDF measurements with custom-made acrylic splints were performed. Six protocols were applied in separate regions of interest (ROI): 1; basal LDF, 2; LDF with thermoprobe 42 °C, 3; LDF with thermoprobe 45 °C, 4; LDF with thermoprobe 42 °C and MN, 5; LDF with thermoprobe 45 °C and MN and 6; LDF with MN.

RESULTS

Intra-individual reproducibility was assessed by the within-subject coefficient of variation (wCV) and the intraclass correlation coefficient (ICC). Basal LDF measurements demonstrated high reproducibility with wCV 11.1 in 2 min and 10.3 in 5 min. ICC was 0.9 and 0.92. wCV after heat and MN was 4.9-10.3 and ICC 0.82-0.93. The topically applied MN yielded increased blood flow.

CONCLUSION

This is the first study evaluating the reproducibility of basal LDF compared to single or multiple vasodilatory stimuli in gingiva. Multiple collector fibers probe and stabilizing acrylic splints are recommended. Vasodilatory stimulation showed a tendency toward higher reproducibility. Furthermore, MN yields vasodilation in gingiva.

摘要

目的

激光多普勒血流仪(LDF)是一种用于评估组织血流的非侵入性技术,但提高其可重复性将有助于纵向研究。本研究的目的是在六个相互关联的方案中,评估激光多普勒血流仪(LDF)在静息状态、局部温度升高时以及使用血管扩张剂烟酸甲酯(MN)的情况下,对牙龈微血管床血供测量的日间可重复性。

方法

纳入10名健康志愿者。使用定制的丙烯酸夹板进行周间LDF测量。在不同的感兴趣区域(ROI)应用六个方案:1;基础LDF,2;使用42°C热探头的LDF,3;使用45°C热探头的LDF,4;使用42°C热探头和MN的LDF,5;使用45°C热探头和MN的LDF,6;使用MN的LDF。

结果

通过受试者内变异系数(wCV)和组内相关系数(ICC)评估个体内可重复性。基础LDF测量显示出高可重复性,2分钟时wCV为11.1,5分钟时为10.3。ICC分别为0.9和0.92。加热和使用MN后的wCV为4.9 - 10.3,ICC为0.82 - 0.93。局部应用MN使血流增加。

结论

这是第一项评估牙龈基础LDF与单一或多种血管舒张刺激相比的可重复性的研究。推荐使用多采集光纤探头和稳定的丙烯酸夹板。血管舒张刺激显示出更高可重复性的趋势。此外,MN可使牙龈血管舒张。