Westmead Hospital, Sydney, New South Wales, Australia
Department of Paediatrics and Child Health, Sydney University, Sydney, New South Wales, Australia.
Arch Dis Child Fetal Neonatal Ed. 2024 Aug 16;109(5):535-541. doi: 10.1136/archdischild-2023-326256.
Respiratory function monitors (RFMs) have been used extensively in manikin and infant studies yet have not become the standard of training. We report the outcomes of a new portable, lightweight RFM, the Juno, designed to show mask leak and deflation tidal volume to assist in positive pressure ventilation (PPV) competency training using manikins.
Two leak-free manikins (preterm and term) were used. Participants provided PPV to manikins using two randomised devices, self-inflating bag (SIB) and T-piece resuscitator (TPR), with Juno display initially blinded then unblinded in four 90 s paired sequences, aiming for adequate chest wall rise and target minimal mask leak with appropriate target delivered volume when using the monitor.
49 experienced neonatal staff delivered 15 569 inflations to the term manikin and 14 580 inflations to the preterm. Comparing blinded to unblinded RFM display, there were significant reductions in all groups in the number of inflations out of target range volumes (preterm: SIB 22.6-6.6%, TPR 7.1-4.2% and term: SIB 54.8-37.8%, TPR 67.2-63.8%). The percentage of mask leak inflations >60% was reduced in preterm: SIB 20.7-7.2%, TPR 23.4-7.4% and in term: SIB 8.7-3.6%, TPR 23.5-6.2%).
Using the Juno monitor during simulated resuscitation significantly improved mask leak and delivered ventilation among otherwise experienced staff using preterm and term manikins. The Juno is a novel RFM that may assist in teaching and self-assessment of resuscitation PPV technique.
呼吸功能监测器(RFMs)已广泛应用于模拟人和婴儿研究中,但尚未成为培训标准。我们报告了一种新的便携式、轻量级 RFM——Juno 的结果,该设备旨在显示面罩泄漏和呼气潮气量,以协助使用模拟人进行正压通气(PPV)能力培训。
使用两个无泄漏模拟人(早产儿和足月儿)。参与者使用两种随机设备(自充气球(SIB)和 T 型复苏器(TPR))向模拟人提供 PPV,Juno 显示器最初处于盲态,然后在四个 90 秒的配对序列中处于非盲态,旨在实现适当的胸壁抬起,并在使用监测器时使用适当的目标输送量实现最小的面罩泄漏。
49 名经验丰富的新生儿科工作人员向足月儿模拟人输送了 15569 次充气,向早产儿模拟人输送了 14580 次充气。与盲态相比,在所有组中,超出目标范围体积的充气次数均显著减少(早产儿:SIB 22.6-6.6%,TPR 7.1-4.2%;足月儿:SIB 54.8-37.8%,TPR 67.2-63.8%)。早产儿 SIB(20.7-7.2%)和 TPR(23.4-7.4%)面罩泄漏充气百分比和足月儿 SIB(8.7-3.6%)和 TPR(23.5-6.2%)面罩泄漏充气百分比均降低。
在模拟复苏过程中使用 Juno 监测器可显著改善面罩泄漏,并提高使用早产儿和足月儿模拟人的经验丰富的工作人员的输送通气。Juno 是一种新型 RFM,可能有助于教授和自我评估复苏 PPV 技术。