Hänni Olivia, Ruby Lisa, Paverd Catherine, Frauenfelder Thomas, Rominger Marga B, Martin Alexander
Faculty of Medicine, University of Zurich, Dekanat Pestalozzistrasse 3, 8032 Zurich, Switzerland.
Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Rämistrasse 100, 8091 Zurich, Switzerland.
Diagnostics (Basel). 2024 Jan 26;14(3):271. doi: 10.3390/diagnostics14030271.
There have been studies showing attenuation imaging (ATI) with ultrasound as an approach to diagnose liver diseases such as steatosis or cirrhosis. So far, this technique has only been used on a convex probe. The goal of the study was to investigate the feasibility of ATI measurements using the linear array on a canon Aplio i800 scanner on certified phantoms. Three certified liver tissue attenuation phantoms were measured in five different positions using a linear probe. The effects of positioning and depth were explored and compared. The values were compared to the certified expected value for each phantom as well as the different measurement values for each measurement position. The ATI measurements on phantoms showed significant effect for the different probe positions and region of interest (ROI) depths. Values taken in the center with the probe perpendicular to the phantom were closest to certified values. Median values at 2.5-4.5 cm depth for phantoms 1 and 2 and 0.5-2.5 cm for phantom 3 were comparable with certified values. Measurements taken at a depth greater than 6 cm in any position were the least representative of the certified values (-value < 0.01) and had the widest range throughout the different sessions. ATI measurements can be performed with the linear probe in phantoms; however, careful consideration should be given to depth dependency, as it can significantly affect measurement values. Remaining measurements at various depths within the 0.5-6.0 cm range showed deviation from the certified values of approximately 25%.
已有研究表明,使用超声衰减成像(ATI)作为诊断脂肪肝或肝硬化等肝脏疾病的一种方法。到目前为止,这项技术仅用于凸阵探头。本研究的目的是探讨在佳能Aplio i800扫描仪上使用线性阵列对认证体模进行ATI测量的可行性。使用线性探头在三个认证的肝脏组织衰减体模的五个不同位置进行测量。探讨并比较了定位和深度的影响。将测量值与每个体模的认证预期值以及每个测量位置的不同测量值进行比较。对体模的ATI测量显示,不同的探头位置和感兴趣区域(ROI)深度有显著影响。探头垂直于体模在中心位置取得的值最接近认证值。体模1和2在2.5 - 4.5厘米深度的中值以及体模3在0.5 - 2.5厘米深度的中值与认证值相当。在任何位置深度大于6厘米处进行的测量最不能代表认证值(p值<0.01),并且在不同测量环节中范围最广。可以使用线性探头在体模中进行ATI测量;然而,应仔细考虑深度依赖性,因为它会显著影响测量值。在0.5 - 6.0厘米范围内不同深度的其余测量值显示与认证值的偏差约为25%。