Evaluation of an apparatus for continuous monitoring of gas exchange in mechanically ventilated patients.

作者信息

Carlsson M, Forsberg E, Thörne A, Nordenström J, Hedenstierna G

出版信息

Int J Clin Monit Comput. 1985;1(4):211-20. doi: 10.1007/BF01720185.

DOI:10.1007/BF01720185
PMID:3836285
Abstract

Indirect calorimetry has been integrated with a commercially available ventilator, permitting continuous monitoring of energy expenditure and respiratory quotient. This equipment, the ventilator Erica, the Engström metabolic computer EMC and the carbon dioxide analyzer Eliza (Gambro Engström AB, Sweden) was evaluated in laboratory tests and in critically ill patients. In the laboratory evaluation, the variability in FIO2 was less than 0.05% at FIO2 less than 0.47. The linearity error of the oxygen sensor and the CO2 analyzer was less than 0.06%. The variability of the oxygen sensor was 0.01% O2 and of the CO2 analyzer 0.04% CO2. The CO2 analyzer was sensitive to oxygen concentration and underestimated CO2 concentration by 3.3% at FIO2 = 0.5 if calibrated on room air. The pneumotachometer was oxygen dependent and overestimated inspired volume by 5.5% at 100% O2, if calibrated on room air, and was negligibly affected by changes in minute volume or PEEP within the physiological range for adults. In the patient evaluation, 15 measurements of gas exchange were performed in 9 critically ill patients and the results were compared with data obtained simultaneously by the Scholander technique or mass spectrometry. The standard deviation of the mean difference in comparison with the reference methods was 9.4% for O2 uptake, 6.6% for CO2 elimination, 12.5% for respiratory quotient and 3.9% for minute volume. Paired t-test analysis showed no systematic difference between the various methods. It is concluded that the EMC has an accuracy sufficient for clinical use and represents a commercially available technique of potential clinical value.

摘要

相似文献

1
Evaluation of an apparatus for continuous monitoring of gas exchange in mechanically ventilated patients.
Int J Clin Monit Comput. 1985;1(4):211-20. doi: 10.1007/BF01720185.
2
[Evaluation of the Engstrom metabolic computer].
Ann Fr Anesth Reanim. 1987;6(2):129-32. doi: 10.1016/s0750-7658(87)80120-x.
3
A new device for continuous measurement of gas exchange during artificial ventilation.一种用于人工通气期间连续测量气体交换的新设备。
Crit Care Med. 1980 Dec;8(12):705-9. doi: 10.1097/00003246-198012000-00001.
4
Accuracy of an indirect calorimeter for mechanically ventilated infants and children: the influence of low rates of gas exchange and varying FIO2.用于机械通气婴幼儿的间接热量计的准确性:低气体交换率和不同吸入氧浓度的影响。
Crit Care Med. 2000 Aug;28(8):3014-8. doi: 10.1097/00003246-200008000-00056.
5
Effect of FiO in the measurement of VO and VCO using the E-COXV metabolic monitor.使用 E-COXV 代谢监测仪时 FiO 对 VO 和 VCO 测量的影响。
Med Intensiva. 2017 Nov;41(8):461-467. doi: 10.1016/j.medin.2016.12.002. Epub 2017 Mar 8.
6
A ventilator with an integrated gas-exchange monitoring function.具有集成气体交换监测功能的呼吸机。
Crit Care Med. 1994 May;22(5):864-71. doi: 10.1097/00003246-199405000-00025.
7
Methods to validate the accuracy of an indirect calorimeter in the in-vitro setting.在体外环境中验证间接热量计准确性的方法。
Clin Nutr ESPEN. 2017 Dec;22:71-75. doi: 10.1016/j.clnesp.2017.08.009. Epub 2017 Aug 30.
8
Respiratory deadspace measurements in neonates during extracorporeal membrane oxygenation.体外膜肺氧合期间新生儿呼吸死腔测量
Crit Care Med. 1993 Dec;21(12):1895-900. doi: 10.1097/00003246-199312000-00017.
9
Evaluation of methods for indirect calorimetry with a ventilated lung model.使用通气肺模型评估间接测热法的方法
Intensive Care Med. 1989;15(3):196-202. doi: 10.1007/BF01058573.
10
Comparison of metabolic monitors in critically ill, ventilated patients.危重症机械通气患者代谢监测仪的比较
Nutrition. 2006 Nov-Dec;22(11-12):1077-86. doi: 10.1016/j.nut.2006.06.007. Epub 2006 Sep 14.

引用本文的文献

1
Energy expenditure and outcome in patients with multiple organ failure following abdominal surgery.腹部手术后多器官功能衰竭患者的能量消耗与预后
Intensive Care Med. 1991;17(7):403-9. doi: 10.1007/BF01720678.