两种用于儿童和青少年视力筛查的不同自动验光仪。

Two different autorefractors for vision screening in children and adolescents.

作者信息

Chen Xiao-Jun, Liu Lin-Jie, Sun Bing, Jiang Dan-Dan, Zuo Shu-Shu, Wang Yan-Hui, Zhang Chun-Hua, Chen Yan-Yan

机构信息

School of Optometry and Ophthalmology, Wenzhou Medical University, Wenzhou 325027, Zhejiang Province, China.

National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou 325027, Zhejiang Province, China.

出版信息

Int J Ophthalmol. 2024 Feb 18;17(2):331-338. doi: 10.18240/ijo.2024.02.16. eCollection 2024.

Abstract

AIM

To compare the consistency of two autorefractors (Tianle RM-9000 and Topcon KR-800) for school-age myopia children, and to provide a basis for largescale data analysis and comparison.

METHODS

The refractive error in 909 subjects (age 4-18y) were measured using both autorefractors without cycloplegia. The data were analyzed using Fourier decomposition and the correlation coefficients, intraclass correlation coefficients (ICC), and Bland-Altman limits of agreement (LoA) for each parameter were calculated.

RESULTS

There was a strong correlation between the spherical equivalent (SE), sphere diopter (DS), and cylinder diopter (DC) readings of the Tianle RM-9000 and those of the Topcon KR-800, with correlation coefficient values of 0.98, 0.98 and 0.83 and ICC values of 0.99, 0.99 and 0.93, respectively. However, the correlation coefficients and ICC values of J and J were unreliable (=-0.004, -0.034; both ICC<0.10). Bland-Altman analysis revealed that SE, DS, and DC measured by the Tianle RM-9000 were significantly biased toward myopia compared with the Topcon KR-800, and the mean differences were -0.072, -0.026, -0.091 D, respectively (all <0.01). The minimum absolute value of the difference within the 95% LoA for SE, DS, and DC was 0.63 D, 0.50 D, 0.62 D, respectively; all these values were in the clinically acceptable range. For J and J, the mean differences were close to zero (=0.43, 0.84); however, the 95% LoA were relatively wide (J SD: 0.53; 95%CI: -1.00, 1.10; J SD: 0.52; 95%CI: -1.00, 1.00).

CONCLUSION

The two autorefractors are consistent with each other, as the differences in SE, DS, and DC were within the clinically acceptable range. Readers can compare the data measured by either device in different studies and use the two devices in the same study to generate a dataset that can be analyzed together. However, the J and J vectors are unreliable and should not be used to assess astigmatism.

摘要

目的

比较两种自动验光仪(天乐RM - 9000和拓普康KR - 800)在学龄期近视儿童中的一致性,为大规模数据分析和比较提供依据。

方法

使用两种自动验光仪在未散瞳的情况下测量909名受试者(年龄4 - 18岁)的屈光不正。采用傅里叶分解分析数据,并计算各参数的相关系数、组内相关系数(ICC)和布兰德 - 奥特曼一致性界限(LoA)。

结果

天乐RM - 9000和拓普康KR - 800的等效球镜度(SE)、球镜度数(DS)和柱镜度数(DC)读数之间存在强相关性,相关系数分别为0.98、0.98和0.83,ICC值分别为0.99、0.99和0.93。然而,J0和J45的相关系数和ICC值不可靠(r = -0.004,-0.034;ICC均<0.10)。布兰德 - 奥特曼分析显示,与拓普康KR - 800相比,天乐RM - 9000测量的SE、DS和DC明显偏向近视,平均差异分别为-0.072、-0.026、-0.091 D(均<0.01)。SE、DS和DC在95% LoA内差异的最小绝对值分别为0.63 D、0.50 D、0.62 D;所有这些值均在临床可接受范围内。对于J0和J45,平均差异接近零(r = 0.43,0.84);然而,95% LoA相对较宽(J0 SD:0.53;95%CI:-1.00,1.10;J45 SD:0.52;95%CI:-1.00,1.00)。

结论

两种自动验光仪相互之间具有一致性,因为SE、DS和DC的差异在临床可接受范围内。读者可以在不同研究中比较任一设备测量的数据,并在同一研究中使用这两种设备生成可一起分析的数据集。然而,J0和J45向量不可靠,不应将其用于评估散光。

相似文献

引用本文的文献

1
Autorefraction versus subjective refraction in high astigmatism.高度散光患者的自动验光与主观验光比较
Int J Ophthalmol. 2025 Mar 18;18(3):504-509. doi: 10.18240/ijo.2025.03.18. eCollection 2025.

本文引用的文献

5
Myopia.近视。
Nat Rev Dis Primers. 2020 Dec 17;6(1):99. doi: 10.1038/s41572-020-00231-4.