To evaluate the clinical applicability of a semiautomatic radiological tool for scalar translocation detection.
Retrospective study.
Tertiary care academic center.
We included 104 patients implanted with 116 HiFocus Mid-Scala electrode arrays between January 2013 and September 2016.
Cochlear implantation.
The tool's scalar position assessments were compared with manual ones by calculating intraclass coefficient (ICC) for individual contacts and sensitivity and specificity for translocation detection of the whole array. In addition, ICC was calculated for diameters A and B, ratio A/B, and angular insertion depth (AID).
Nine-one percent of cases could be processed, which took 5 to 10 minutes per case. Comparison of manual and semiautomatic scalar position showed for individual contacts an ICC of 0.89 and for the whole array a sensitivity of 97% and a specificity of 96%. ICCs for A, B, and A/B were 0.82, 0.74, and 0.39 respectively. For AID, ICC of each of the 16 contacts was 0.95 or higher.
The semiautomatic radiological tool could analyze most cases and showed good to excellent agreement with manual assessments for translocation detection, diameter A, diameter B, and AID. The variability between semiautomatic and manual measurements is comparable to interobserver variability, indicating that clinical implementation of the tool is feasible.
评估一种半自动放射学工具在检测标度移位方面的临床适用性。
回顾性研究。
三级保健学术中心。
我们纳入了 2013 年 1 月至 2016 年 9 月期间植入 116 个 HiFocus Mid-Scala 电极阵列的 104 名患者。
人工耳蜗植入。
通过计算个别触点的组内相关系数(ICC)以及整个数组的移位检测的灵敏度和特异性,比较工具的标度位置评估与手动评估。此外,还计算了直径 A 和 B、A/B 比值和角插入深度(AID)的 ICC。
91%的病例可以进行处理,每个病例需要 5 到 10 分钟。手动和半自动标度位置比较显示,个别触点的 ICC 为 0.89,整个数组的灵敏度为 97%,特异性为 96%。A、B 和 A/B 的 ICC 分别为 0.82、0.74 和 0.39。对于 AID,每个触点的 ICC 均在 0.95 或更高。
半自动放射学工具可以分析大多数病例,并且在检测移位、直径 A、直径 B 和 AID 方面与手动评估具有良好到极好的一致性。半自动和手动测量之间的可变性与观察者之间的可变性相当,表明该工具的临床应用是可行的。