Donovan R M, Goldstein E
J Histochem Cytochem. 1985 Jun;33(6):551-6. doi: 10.1177/33.6.3839005.
A rapid, semiautomated cytophotometry system for quantitative histochemistry and cytochemistry was constructed. The system consists of a Fairchild charge coupled device (CCD) image camera, a Zeiss Universal microscope, a Datacube analog to digital converter, and a digital Equipment Corporation LSI 11/23 computer operating under RT-11. Computer programs were written in FORTRAN and the MACRO assembly language for the acquisition of data from the CCD device. These data were then used for image segmentation, image display, and calculation of total optical density, perimeter, cell area, and several shape features. The reproducibility of measurement made with the CCD-based cytophotometry system was tested by repeated measurements. The coefficient of variation was estimated to be 1.7% for total optical density and 0.9% for cell area. The CCD-based cytophotometry system was further evaluated by comparing results with measurements made on the same cells with a scanning stage cytophotometer using the HIDACSYS computer programs. Correlation coefficients of 0.96 for total optical density and 0.91 for cell area were obtained between the two systems. We conclude that the high-speed, dimensional stability, small size, and linearity of the CCD-based cytophotometry system will make it useful for quantitative histochemistry and cytochemistry.
构建了一种用于定量组织化学和细胞化学的快速半自动细胞光度测量系统。该系统由仙童电荷耦合器件(CCD)图像相机、蔡司万能显微镜、数据立方模数转换器以及在RT - 11系统下运行的数字设备公司LSI 11/23计算机组成。用FORTRAN语言和宏汇编语言编写了计算机程序,用于从CCD设备采集数据。这些数据随后用于图像分割、图像显示以及总光密度、周长、细胞面积和几个形状特征的计算。通过重复测量测试了基于CCD的细胞光度测量系统测量结果的可重复性。总光密度的变异系数估计为1.7%,细胞面积的变异系数估计为0.9%。通过将结果与使用HIDACSYS计算机程序的扫描台细胞光度计对相同细胞进行的测量结果进行比较,进一步评估了基于CCD的细胞光度测量系统。两个系统之间总光密度的相关系数为0.96,细胞面积的相关系数为0.91。我们得出结论,基于CCD的细胞光度测量系统的高速、尺寸稳定性、小尺寸和线性度使其在定量组织化学和细胞化学中很有用。