Department of Haematology, The Royal London Hospital, London, UK.
School of Pharmacy and Biomedical Sciences, University of Portsmouth, Portsmouth, UK.
Scand J Clin Lab Invest. 2024 Apr;84(2):91-96. doi: 10.1080/00365513.2024.2321590. Epub 2024 Feb 29.
The Immature Platelet Fraction (IPF) is an indicator of thrombopoiesis which is a useful parameter in thrombocytopenia. It demonstrates compensatory mechanisms in production of platelets, but currently not implemented in routine clinical practice. The aim of this study was to establish the reproducibility and stability of IPF, for both percentage (%-IPF) and absolute (A-IPF) measurements. A total of 71 samples, of which 45 for reproducibility and 26 for stability analysis, were assayed for full blood count using the Sysmex XN-10 analyser at room temperature (RT:19-25 °C). For reproducibility analysis, IPF measurements were analysed 11 times by different appraisers using the same sample, while for stability analysis, IPF was measured over fourteen hourly-intervals up to 24 h ( = 21) and then separately extended beyond the point of stability to 72 h ( = 5).
Reproducibility analysis of %-IPF and A-IPF ( = 45) showed very reliable results, with the range of mean CV% values between 1.25-8.90% and 1.70-9.96%, respectively. On the other hand, overall, stability analysis of %-IPF and A-IPF ( = 21) at RT over 24 h showed reliable results, with pooled mean CV% values of 1.32% and 1.43%, respectively, with no significant difference between %-IPF and A-IPF ( = 0.767 and = 0.821). All %-IPF and A-IPF values had exceeded the set acceptance criterion of stability (CV% ≥ 10.0%) before 72 h.
Overall, %-IPF and A-IPF reproducibility and storage at RT for 24 h predominantly demonstrates the suitability of their usage for testing on the Sysmex XN-series analysers.
未成熟血小板分数(IPF)是血小板生成的一个指标,在血小板减少症中是一个有用的参数。它显示了血小板生成的代偿机制,但目前尚未在常规临床实践中实施。本研究的目的是建立 IPF 的可重复性和稳定性,包括百分比(%-IPF)和绝对值(A-IPF)测量。共分析了 71 个样本,其中 45 个用于重复性分析,26 个用于稳定性分析,使用 Sysmex XN-10 分析仪在室温(19-25°C)下进行全血细胞计数。为了进行重复性分析,不同评估者使用同一样本进行了 11 次 IPF 测量,而对于稳定性分析,在 14 个小时间隔内测量了 IPF,最长可达 24 小时( = 21),然后分别延长至稳定性点以外 72 小时( = 5)。
重复性分析的 %-IPF 和 A-IPF( = 45)结果非常可靠,平均 CV%值范围分别为 1.25-8.90%和 1.70-9.96%。另一方面,在室温下 24 小时内,总体稳定性分析的 %-IPF 和 A-IPF( = 21)结果可靠,平均 CV%值分别为 1.32%和 1.43%,两者之间无显著差异( = 0.767 和 = 0.821)。在 72 小时之前,所有 %-IPF 和 A-IPF 值都超过了稳定性设定的可接受标准(CV%≥10.0%)。
总体而言,%-IPF 和 A-IPF 的重复性和在室温下储存 24 小时主要表明它们适用于在 Sysmex XN 系列分析仪上进行测试。