Suppr超能文献

基于阻抗的方法评估在人工耳蜗植入过程中监测电极阵列插入的效果。

Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation.

机构信息

Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria.

MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria.

出版信息

Eur Arch Otorhinolaryngol. 2024 Aug;281(8):4121-4131. doi: 10.1007/s00405-024-08584-2. Epub 2024 Apr 2.

Abstract

PURPOSE

Cochlear implantation is a prevalent remedy for severe-to-profound hearing loss. Optimising outcomes and hearing preservation, and minimising insertion trauma, require precise electrode placement. Objective monitoring during the insertion process can provide valuable insights and enhance surgical precision. This study assesses the feasibility and performance of an impedance-based method for monitoring electrode insertion, compared to the surgeon's feedback.

METHODS

The study utilised the Insertion Monitoring Tool (IMT) research software, allowing for real-time measurement of impedance and evoked compound action potential (eCAP) during electrode insertion in 20 patient implantations. This enabled an impedance-based method to continuously assess the status of each electrode during the insertion process. The feasibility and performance was evaluated and compared to the surgeon's feedback approach. eCAP measurements focused merely on feasibility without searching specific responses.

RESULTS

The IMT demonstrated feasibility in measuring real-time impedances and eCAP during the insertion of the electrode array. The impedance-based method exhibited potential for accurately monitoring the insertion depth with a high success rate. However, further development is needed to improve the number of usable contacts.

CONCLUSIONS

Objective monitoring with the impedance-based method shows promise as a valuable tool to enhance the precision of cochlear implant electrode insertion respecting insertion distance estimation. The IMT research software proved feasible in recording real-time impedances and eCAP during electrode insertion. While this impedance-based method exhibits high success rates, further improvements are required to optimise the number of usable contacts. This study highlights the potential of objective monitoring techniques to enhance cochlear implantation outcomes.

摘要

目的

人工耳蜗植入是治疗重度至极重度听力损失的常见方法。为了优化结果和听力保护,减少插入创伤,需要精确放置电极。在插入过程中进行客观监测可以提供有价值的见解并提高手术精度。本研究评估了基于阻抗的方法在监测电极插入方面的可行性和性能,与外科医生的反馈进行比较。

方法

该研究使用了插入监测工具(IMT)研究软件,允许在 20 例患者植入过程中实时测量电极插入过程中的阻抗和诱发复合动作电位(eCAP)。这使得基于阻抗的方法能够在插入过程中连续评估每个电极的状态。评估并比较了该方法的可行性和性能与外科医生的反馈方法。eCAP 测量仅关注可行性,而不寻找特定反应。

结果

IMT 在测量电极阵列插入过程中的实时阻抗和 eCAP 方面表现出可行性。基于阻抗的方法在准确监测插入深度方面具有很高的成功率,但需要进一步开发以提高可用触点的数量。

结论

基于阻抗的客观监测方法具有作为一种有价值的工具的潜力,可以增强人工耳蜗植入电极插入的精度,尊重插入距离的估计。IMT 研究软件在记录电极插入过程中的实时阻抗和 eCAP 方面表现出可行性。虽然这种基于阻抗的方法具有很高的成功率,但需要进一步改进以优化可用触点的数量。本研究强调了客观监测技术在提高人工耳蜗植入效果方面的潜力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0c4a/11266372/ebcb61602014/405_2024_8584_Fig1_HTML.jpg