消毒、调磨和热老化对氧化锆基台固位力的影响:一项实验室研究。

Effects of sterilization, conditioning, and thermal aging on the retention of zirconia hybrid abutments: A laboratory study.

机构信息

Department of Prosthodontics, Propaedeutics and Dental Materials, School of Dentistry, Christian-Albrechts University, Kiel, Germany.

Department of Prosthodontics, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia.

出版信息

Clin Implant Dent Relat Res. 2024 Aug;26(4):679-687. doi: 10.1111/cid.13332. Epub 2024 Apr 29.

Abstract

OBJECTIVE

To investigate the effects of sterilization, conditioning method, and thermal aging on the retentive strength of two-piece zirconia abutments.

MATERIALS AND METHODS

In total, 128 stock zirconia abutments were divided into four groups (n = 32) according to the conditioning parameters: (A) air-abrasion using 50 μm alumina particles/1.0 bar, (B) 50 μm/2.0 bar, (C) 100 μm/1.0 bar, and (D) 100 μm/2.0 bar. All abutments were bonded onto titanium bases using DTK adhesive resin and stored in water bath (37°C) for 72 h. Each group was subdivided into two subgroups (n = 16), group 1 was disinfected, whereas group 2 followed disinfection and autoclave sterilization. Half of the specimens of each subgroup (n = 8) was directly subjected to the axial retention test (groups N), while the other half was first subjected to 150 days of thermocycling followed by retention test (groups T). Statistical analysis was performed with three-way ANOVA, additional statistical analysis was performed by using separate one-way ANOVAs followed by the Tukey's post-hoc test for post hoc pairwise comparisons among groups.

RESULTS

The highest median retention strength was recorded for group B2N (1390 N), whereas the lowest strength was recorded for group C1T (688 N). No significant interaction (p ≥ 0.05) was detected between the different variables; conditioning method, sterilization, and the thermal cycling regarding the effect on the resulting retention. However, the sterilization always showed a positive effect. Thermocycling presented an adversely significant effect only in the absence of sterilization (p < 0.05), with the exception of subgroups A. For the sterilized groups, thermocycling had no statistically significant effect on the retention.

CONCLUSION

Steam autoclaving increased the retention of hybrid zirconia abutments. DTK adhesive resin for two-piece zirconia abutments performed well after sterilization and thermocycling.

摘要

目的

研究灭菌、调理方法和热老化对两段式氧化锆基台固位力的影响。

材料与方法

将 128 个库存氧化锆基台按调理参数分为 4 组(n=32):(A)使用 50μm 氧化铝颗粒/1.0bar 进行气喷砂,(B)50μm/2.0bar,(C)100μm/1.0bar,和(D)100μm/2.0bar。所有基台均使用 DTK 胶粘剂树脂黏合到钛基底上,并在水浴(37°C)中储存 72h。每组再分为 2 个亚组(n=16),亚组 1 经消毒,亚组 2 随后进行消毒和高压蒸汽灭菌。每个亚组的一半样本(n=8)直接进行轴向固位力测试(组 N),另一半样本先进行 150 天热循环,然后进行固位力测试(组 T)。采用三因素方差分析进行统计学分析,对具有统计学意义的变量(p<0.05)采用单独的单向方差分析,然后进行 Tukey 事后检验进行组间两两比较。

结果

组 B2N 的中位固位力最高(1390N),组 C1T 的最低(688N)。不同变量(调理方法、灭菌和热循环)对固位力的影响之间未检测到显著交互作用(p≥0.05)。然而,灭菌始终显示出积极的影响。仅在未灭菌的情况下,热循环才表现出显著的不利影响(p<0.05),A 亚组除外。对于已灭菌的组,热循环对固位力没有统计学上的显著影响。

结论

蒸汽高压灭菌增加了混合氧化锆基台的固位力。两段式氧化锆基台用 DTK 胶粘剂树脂经灭菌和热循环后性能良好。