Clin Oral Investig. 2019 Jul;23(7):3133-3137. doi: 10.1007/s00784-019-02926-9. Epub 2019 May 10.
Aim was to assess the influence of a standard autoclaving protocol on the retention of zirconia abutments glued on titanium bases in two-piece implant abutments.
Twenty-four CAD/CAM-generated zirconia abutments were adhesively cemented on prefabricated titanium bases. Before mechanical and thermal aging, the specimens were divided into two groups. Group 1 was subjected to a standard steam autoclaving protocol and Group 2 remained untreated (control). The tensile strength in all specimens was evaluated by a standardized pull-off test limited to a maximum force of 1000 N.
Eleven samples in both groups could be subjected to pull-off testing. Ten samples in Group 1 and three samples in Group 2 failed, while all others reached the maximum pull-off force. This difference was statistically significant. The mean retention values for the failed samples were 694.53 ± 369.10 N in Group 1 and 890.78 ± 25.90 N in Group 2. This difference was not statistically significant.
A standard autoclaving protocol does not reduce detachment force of two-piece zirconia abutments.
Clinical sterilization processes as recommended by regulatory authorities seem to be harmless to the structural integrity of two-piece zirconia implant abutments, at least with regard to the retention of the components.
评估标准高压蒸汽灭菌方案对 2 件式种植体基台上粘合的氧化锆基台保留的影响。
将 24 个 CAD/CAM 生成的氧化锆基台粘接到预制钛基底上。在进行机械和热老化之前,将标本分为两组。第 1 组接受标准蒸汽高压灭菌方案,第 2 组未处理(对照组)。通过标准化拉脱试验评估所有标本的拉伸强度,最大拉力限制为 1000 N。
两组中均有 11 个样本可进行拉脱试验。第 1 组中有 10 个样本和第 2 组中有 3 个样本失效,而其余所有样本均达到最大拉脱力。这一差异具有统计学意义。第 1 组失效样本的平均保留值为 694.53 ± 369.10 N,第 2 组为 890.78 ± 25.90 N。这一差异无统计学意义。
标准高压蒸汽灭菌方案不会降低 2 件式氧化锆基台的脱落力。
监管机构推荐的临床灭菌过程似乎对 2 件式氧化锆种植体基台的结构完整性没有危害,至少在组件的保留方面没有危害。