Matsumoto Yukihiro, Azuma Yuichi, Karasawa Yasue, Suzuki Noriyuki
Eye Care Clinic, Saitama, Japan.
Eye Care Clinic Tokyo, Tokyo, Japan.
Clin Ophthalmol. 2024 Apr 25;18:1117-1124. doi: 10.2147/OPTH.S456609. eCollection 2024.
To evaluate postoperative outcomes after implantation of toric intraocular lenses (IOLs) made of high-water-content hydrophobic acrylic material in Japanese patients using a swept-source optical coherence tomography (SS-OCT) biometer integrated with a surgical guiding system.
In this prospective observational study, toric IOL models CNW0T3 to CNW0T9 (Alcon) were implanted in 33 eyes of 33 patients and followed-up for one month. Powers and toric models were determined using an SS-OCT biometer ARGOS Ver 1.5 (Alcon), and the IOLs were aligned using surgical guidance. Differences between planned and actual axis positions at the end of the surgery (misalignment) and rotations from the end of surgery to one month postoperatively were measured. Additionally, postoperative uncorrected visual acuity, refraction, and residual astigmatism were evaluated.
Mean and median misalignments were 2.3° (standard deviation [SD]: 1.6, 95% confidence interval [CI]: 1.7-2.9) and 2°, and those of postoperative rotation were 2.4° (SD: 2.6, 95% CI: 1.5-3.4) and 2°, respectively. Mean postoperative refraction was 0.06 D (SD: 0.62). Prediction errors within ±0.5 and ±1.0 D were 69.7% and 93.9%, respectively. Mean residual astigmatisms were 0.19 D (SD: 0.41), and mean uncorrected visual acuity was 0.00 logMAR (SD: 0.11), and 64% of the eyes scored 20/20 or better.
Implantation of high-water-content hydrophobic acrylic toric IOLs using SS-OCT biometry integrated with a surgical guiding system effectively corrected corneal astigmatism with accurate IOL alignment in Japanese patients.
使用集成手术引导系统的扫频光学相干断层扫描(SS-OCT)生物测量仪,评估日本患者植入高含水量疏水性丙烯酸材料的散光人工晶状体(IOL)后的术后结果。
在这项前瞻性观察研究中,将 toric IOL 型号 CNW0T3 至 CNW0T9(爱尔康)植入 33 例患者的 33 只眼中,并随访 1 个月。使用 SS-OCT 生物测量仪 ARGOS Ver 1.5(爱尔康)确定屈光度和 toric 型号,并使用手术引导使 IOL 对齐。测量手术结束时计划轴位与实际轴位之间的差异(错位)以及从手术结束到术后 1 个月的旋转情况。此外,评估术后未矫正视力、屈光和残余散光。
平均和中位数错位分别为 2.3°(标准差[SD]:1.6,95%置信区间[CI]:1.7 - 2.9)和 2°,术后旋转的平均和中位数分别为 2.4°(SD:2.6,95%CI:1.5 - 3.4)和 2°。术后平均屈光为 0.06 D(SD:0.62)。±0.5 D 和±1.0 D 内的预测误差分别为 69.7%和 93.9%。平均残余散光为 0.19 D(SD:0.41),平均未矫正视力为 0.00 logMAR(SD:0.11),64%的眼睛视力达到 20/20 或更好。
在日本患者中,使用集成手术引导系统的 SS-OCT 生物测量法植入高含水量疏水性丙烯酸散光 IOL 可有效矫正角膜散光,并实现 IOL 的精确对齐。