Evaluation of clarity characteristics in a new hydrophobic acrylic IOL in comparison to commercially available IOLs.

机构信息

John A. Moran Eye Center, University of Utah, Salt Lake City, Utah, USA.

Alcon Vision, LLC, Fort Worth, Texas, USA.

出版信息

J Cataract Refract Surg. 2019 Oct;45(10):1490-1497. doi: 10.1016/j.jcrs.2019.05.017. Epub 2019 Aug 6.

Abstract

PURPOSE

To perform in vitro assessments of surface haze and roughness, subsurface nanoglistenings (SSNGs), and glistenings of a new hydrophobic acrylic intraocular lens (IOL) in comparison to commercially available IOLs.

SETTING

Universities of Utah and Alcalá, and Alcon Vision, LLC.

DESIGN

Experimental study.

METHODS

The IOLs used were Clareon CNA0T0, Tecnis ZCB00 and Tecnis OptiBlue ZCB00V, Eternity W-60, enVista MX60, and Vivinex XY1. Surface haze was assessed by the pixel intensity units (PIU) from cross-sectional slitlamp images, and it was correlated to atomic force microscopy roughness. For SSNGs, the IOLs were artificially aged up to 10 years, followed by Scheimpflug densitometry (computer-compatible tape unit [CCTU]) measurements. For glistenings, the IOLs were immersed in a water bath and subjected to temperature changes before analysis.

RESULTS

The surface haze (n = 10, PIU) was 4.25 ± 0.87 (SD) (CNA0T0), 9.50 ± 1.66 (ZCB00), 39.48 ± 1.97 (ZCB00V), 46.68 ± 3.16 (W-60), 44.70 ± 4.00 (MX60), and 4.42 ± 0.71 (XY1) (P < .001), which showed a strong correlation with surface roughness measurements (R = 0.94, P = .006). The densitometry (n = 10, CCTU at 10 years) was 7.30 ± 1.36 (CNA0T0), 11.88 ± 3.10 (ZCB00), 38.12 ± 2.24 (ZCB00V), 48.13 ± 9.44 (W-60), 20.20 ± 3.84 (MX60), and 6.75 ± 6.66 (XY1) (P < .001), with no significant differences between CNA0T0 vs. ZCB00 and XY1. The glistenings density (n = 30) was the lowest for the CNA0T0, W-60, and MX60 IOLs.

CONCLUSIONS

The new Clareon CNA0T0, which is manufactured from a hydrophobic acrylic material, exhibited among the lowest levels of surface haze and roughness, SSNGs, and glistenings compared with other commercially available hydrophobic acrylic IOLs.

摘要

目的

与市售人工晶状体(IOL)相比,对新型疏水性丙烯酸 IOL 的表面雾化和粗糙度、亚表面纳米划痕(SSNGs)和光泽度进行体外评估。

地点

犹他大学和阿尔卡拉大学,以及爱尔康视觉有限责任公司。

设计

实验研究。

方法

使用的 IOL 有 Clareon CNA0T0、Tecnis ZCB00 和 Tecnis OptiBlue ZCB00V、Eternity W-60、enVista MX60 和 Vivinex XY1。使用横截面裂隙灯图像的像素强度单位(PIU)评估表面雾化,并将其与原子力显微镜粗糙度相关联。对于 SSNGs,将 IOL 人工老化长达 10 年,然后进行 Scheimpflug 密度测量(计算机兼容磁带单元 [CCTU])。对于光泽度,将 IOL 浸入水浴中,然后在分析前进行温度变化。

结果

表面雾化(n=10,PIU)为 4.25±0.87(SD)(CNA0T0)、9.50±1.66(ZCB00)、39.48±1.97(ZCB00V)、46.68±3.16(W-60)、44.70±4.00(MX60)和 4.42±0.71(XY1)(P<0.001),与表面粗糙度测量值具有很强的相关性(R=0.94,P=0.006)。密度测量值(n=10,10 年 CCTU)为 7.30±1.36(CNA0T0)、11.88±3.10(ZCB00)、38.12±2.24(ZCB00V)、48.13±9.44(W-60)、20.20±3.84(MX60)和 6.75±6.66(XY1)(P<0.001),CNA0T0 与 ZCB00 和 XY1 之间无显著差异。CNA0T0、W-60 和 MX60 IOL 的光泽度密度最低。

结论

新型疏水性丙烯酸 Clareon CNA0T0 与其他市售疏水性丙烯酸 IOL 相比,其表面雾化和粗糙度、SSNGs 和光泽度最低。