The Impact of Back Optic Zone Design in Orthokeratology on Visual Performance.

机构信息

Clinical College of Ophthalmology, Tianjin Medical University, Tianjin, China.

Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Eye Hospital, Tianjin, China.

出版信息

Transl Vis Sci Technol. 2024 May 1;13(5):12. doi: 10.1167/tvst.13.5.12.

Abstract

PURPOSE

To evaluate the visual performance in adolescents undergoing orthokeratology (OrthoK) treatment with two different optical zone diameters (OZDs).

METHODS

This randomized, double-masked, self-controlled prospective study was conducted at Tianjin Eye Hospital (Tianjin, China) in June 2022. Thirty-six eligible schoolchildren were enrolled and fitted with corneal refractive therapy lenses with two sizes of OZDs (5 mm [5OZ] and 6 mm [6OZ]). Each participant was randomized to wear the 5OZ in one eye and the 6OZ in the contralateral eye. Subjective visual quality was assessed using visual acuity, refraction, contrast sensitivity function, and visual symptoms, and the objective optical quality was assessed using ocular higher order aberrations (HOAs) and modulation transfer function (MTF).

RESULTS

Thirty-five myopic children completed a 1-month follow-up visit. The 5OZ lens induced significantly smaller treatment zone diameters than the 6OZ lens (P < 0.001). Subjective visual quality did not differ significantly between the two groups. Compared to baseline, aberrations of Z40, coma-like, spherical-like, and total HOAs in both groups increased significantly (P < 0.05). For the 3-mm pupils, spherical aberration in the 5OZ group was significantly higher than that in the 6OZ group (P < 0.05). The MTF value of the 6OZ group was significantly higher than that of 5OZ group for 0.3 and 1.5 cycles per degree for the 3-mm pupils (P = 0.006 and P = 0.026, respectively). However, HOAs or MTF did not differ significantly between the two groups for the 5-mm pupils.

CONCLUSIONS

The difference induced by varying OZD was significant only in the smaller pupil condition. The selection of OZD in OrthoK designs in real-world patient management should be done while considering individual pupil size.

TRANSLATIONAL RELEVANCE

This study revealed that the objective visual quality of small OZD lenses was only slightly affected for the small pupil size.

摘要

目的

评估两种不同光学区直径(OZD)下青少年角膜塑形术(OrthoK)治疗的视觉表现。

方法

本研究为 2022 年 6 月在中国天津眼科医院进行的一项随机、双盲、自身对照前瞻性研究。共纳入 36 名符合条件的学龄儿童,并为他们配备了两种光学区直径(5mm[5OZ]和 6mm[6OZ])的角膜屈光治疗镜片。每个参与者被随机分配到一只眼佩戴 5OZ,另一只眼佩戴 6OZ。通过视力、屈光度、对比敏感度函数和视觉症状评估主观视觉质量,通过眼高阶像差(HOAs)和调制传递函数(MTF)评估客观光学质量。

结果

35 名近视儿童完成了 1 个月的随访。5OZ 镜片的治疗区直径明显小于 6OZ 镜片(P<0.001)。两组间主观视觉质量无显著差异。与基线相比,两组的 Z40 像差、彗差、球差和总 HOAs 均显著增加(P<0.05)。对于 3mm 瞳孔,5OZ 组的球差明显高于 6OZ 组(P<0.05)。对于 3mm 瞳孔,6OZ 组的 MTF 值在 0.3 和 1.5 周期/度时明显高于 5OZ 组(P=0.006 和 P=0.026)。然而,对于 5mm 瞳孔,两组之间的 HOAs 或 MTF 无显著差异。

结论

仅在小瞳孔条件下,OZD 变化引起的差异才具有统计学意义。在现实患者管理中,OrthoK 设计中 OZD 的选择应考虑个体瞳孔大小。