Fan Mingdong, Zhou Zhongxing, Wellinghoff Jarod, McCollough Cynthia H, Yu Lifeng
Mayo Clinic, Department of Radiology, Rochester, Minnesota, United States.
J Med Imaging (Bellingham). 2024 Dec;11(Suppl 1):S12803. doi: 10.1117/1.JMI.11.S1.S12803. Epub 2024 May 24.
We aim to compare the low-contrast detectability of a clinical whole-body photon-counting-detector (PCD)-CT at different scan modes and image types with an energy-integrating-detector (EID)-CT.
We used a channelized Hotelling observer (CHO) previously optimized for quality control purposes. An American College of Radiology CT accreditation phantom was scanned on both PCD-CT and EID-CT with 10 phantom positionings. For PCD-CT, images were generated using two scan modes, standard resolution (SR) and ultra-high-resolution (UHR); two image types, virtual monochromatic images at 70 keV and low-energy threshold (T3D); both filtered-back-projection (FBP) and iterative reconstruction (IR) reconstruction methods; and three reconstruction kernels. For each positioning, three repeated scans were acquired for each scan mode, image type, and CTDIvol of 6, 12, and 24 mGy. For EID-CT, images acquired from scans (10 positionings × 3 repeats × 3 doses) were reconstructed using the closest counterpart FBP and IR kernels. CHO was applied to calculate the index of detectability () on both scanners.
With the smooth Br44 kernel, the of UHR was mostly comparable with that of the SR mode (difference: -11.4% to 8.3%, to 0.956), and the T3D images had a higher (difference: 0.7% to 25.6%) than 70 keV images on PCD-CT. Compared with the EID-CT, UHR-T3D of PCD-CT had non-inferior (difference: -2.7% to 12.9%) with IR and non-superior (difference: 0.8% to 11.2%) with FBP using the Br44 kernel. PCD-CT produced higher than EID-CT by 61.8% to 247.1% with the sharper reconstruction kernels.
The comparison between PCD-CT and EID-CT was significantly influenced by the reconstruction method and kernel. With a smooth kernel that is typically used in low-contrast detection tasks, the PCD-CT demonstrated low-contrast detectability that was comparable to EID-CT with IR and showed no superiority when using FBP. With the use of sharper kernels, the PCD-CT significantly outperformed EID-CT in low-contrast detectability.
我们旨在比较临床全身光子计数探测器(PCD)CT在不同扫描模式和图像类型下与能量积分探测器(EID)CT的低对比度可探测性。
我们使用了先前为质量控制目的而优化的通道化霍特林观察者(CHO)。在PCD-CT和EID-CT上对美国放射学会CT认证体模进行扫描,体模有10种定位。对于PCD-CT,使用两种扫描模式生成图像,即标准分辨率(SR)和超高分辨率(UHR);两种图像类型,70 keV的虚拟单色图像和低能量阈值(T3D);滤波反投影(FBP)和迭代重建(IR)两种重建方法;以及三种重建核。对于每种定位,针对每种扫描模式、图像类型以及6、12和24 mGy的CTDIvol,采集三次重复扫描。对于EID-CT,使用最接近的对应FBP和IR核重建从扫描中获取的图像(10种定位×3次重复×3种剂量)。应用CHO计算两台扫描仪上的可探测性指数()。
使用平滑的Br44核时,PCD-CT上UHR的大多与SR模式相当(差异:-11.4%至8.3%,至0.956),且T3D图像的高于70 keV图像(差异:0.7%至25.6%)。与EID-CT相比,使用Br44核时,PCD-CT的UHR-T3D在IR下具有非劣效的(差异:-2.7%至12.9%),在FBP下具有非优效的(差异:0.8%至11.2%)。使用更锐利的重建核时,PCD-CT产生的比EID-CT高61.8%至247.1%。
PCD-CT和EID-CT之间的比较受重建方法和核的显著影响。使用通常用于低对比度检测任务的平滑核时,PCD-CT的低对比度可探测性与IR下的EID-CT相当,使用FBP时无优势。使用更锐利的核时,PCD-CT在低对比度可探测性方面明显优于EID-CT。