Sci Rep. 2024 Jul 2;14(1):15206. doi: 10.1038/s41598-024-66150-6.
The study investigated the effects of temperature and centrifugation time on the efficacy of removing uncured resin from 3D-printed clear aligners. Using a photo-polymerizable polyurethane resin (Tera Harz TC-85, Graphy Inc., Seoul, Korea), aligners were printed and subjected to cleaning processes using isopropyl alcohol (IPA) or centrifugation (g-force 27.95g) at room temperature (RT, 23 °C) and high temperature (HT, 55 °C) for 2, 4, and 6 min. The control group received no treatment (NT). Cleaning efficiency was assessed through rheological analysis, weight measurement, transparency evaluation, SEM imaging, 3D geometry evaluation, stress relaxation, and cell viability tests. Results showed increased temperature and longer centrifugation times significantly reduced aligner viscosity, weight (P < 0.05), and transmittance. IPA-cleaned aligners exhibited significantly lower transparency and rougher surfaces in SEM images. All groups met ISO biocompatibility standards in cytotoxicity tests. The NT group had higher root mean square (RMS) values, indicating greater deviation from the original design. Stress relaxation tests revealed over 95% recovery in all groups after 60 min. The findings suggest that a 2-min HT centrifugation process effectively removes uncured resin without significantly impacting the aligners' physical and optical properties, making it a clinically viable option.
本研究旨在探讨温度和离心时间对从 3D 打印透明矫正器中去除未固化树脂效果的影响。使用光聚合性聚氨酯树脂(Tera Harz TC-85,Graphy Inc.,首尔,韩国),打印出矫正器,并在室温(23°C,RT)和高温(55°C)下分别使用异丙醇(IPA)或离心(g-force 27.95g)清洗 2、4 和 6 分钟。对照组不进行任何处理(NT)。通过流变分析、称重、透明度评估、SEM 成像、3D 几何评估、应力松弛和细胞活力测试评估清洗效率。结果表明,升高温度和延长离心时间会显著降低矫正器的粘度、重量(P<0.05)和透光率。用 IPA 清洗的矫正器在 SEM 图像中表现出明显较低的透明度和更粗糙的表面。所有组在细胞毒性测试中均符合 ISO 生物相容性标准。NT 组的均方根(RMS)值较高,表明与原始设计的偏差更大。在 60 分钟后,所有组的应力松弛测试均显示超过 95%的恢复。研究结果表明,2 分钟的高温离心过程可以有效去除未固化树脂,而不会显著影响矫正器的物理和光学性能,因此是一种具有临床应用价值的选择。