Hirai Kenta, Sakano Noriko, Oozawa Susumu, Ousaka Daiki, Kuroko Yosuke, Kasahara Shingo
Department of Pediatrics, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Science, Okayama, Japan.
Department of Cardiovascular Surgery, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Science, Okayama, Japan.
J Cardiol Cases. 2024 Apr 16;30(1):24-28. doi: 10.1016/j.jccase.2024.03.004. eCollection 2024 Jul.
Sudden cardiac arrest during exercise can occur without prior warning signs at rest, highlighting the importance of monitoring for its prevention. To detect the signs of ischemic heart disease, including coronary artery anomalies, ST changes must be detected using three‑lead electrocardiograms (ECGs) corresponding to each region of the three coronary artery branches. We conducted ECG monitoring of five runners during a marathon using a wearable three‑lead ECG device (e-skin ECG; Xenoma Inc., Tokyo, Japan). Data without noise or artifacts were successfully collected for one of five runners during the entire marathon. Within the initial hour of the marathon, poor electrode adhesion to the skin hindered the data collection for the remaining four runners, which resulted in significantly decreased acquisition rate compared with the first hour (86.7 ± 13.4 % to 37.3 ± 36.9 %, = 0.028). Couplets of premature ventricular contractions with clear ECG waveforms in the three leads were detected in one runner during the marathon. Further device improvements are necessary to enable marathon runners to obtain ECGs efficiently without affecting their performance. This study also demonstrated the potential applications of three‑lead wearable ECG monitoring for other short-duration sports and remote home-based cardiac rehabilitation.
This is an initial trial of a three‑lead wearable electrocardiogram (ECG) monitoring device during a full marathon. ECG data were obtained with low noise and artifacts during the first hour of the marathon; however, the data acquisition rate decreased in the middle and late stages owing to poor electrode adhesion. This study demonstrated the possibility of applying wearable ECG monitoring during short-term exercise and cardiac rehabilitation to detect warning signs and prevent sudden cardiac arrest.
运动期间的心脏骤停可能在休息时没有先前预警信号的情况下发生,这凸显了监测以预防心脏骤停的重要性。为了检测包括冠状动脉异常在内的缺血性心脏病迹象,必须使用对应于三支冠状动脉分支各区域的三导联心电图(ECG)来检测ST段变化。我们在马拉松比赛期间使用可穿戴三导联ECG设备(电子皮肤ECG;日本东京Xenoma公司)对五名跑步者进行了ECG监测。在整个马拉松比赛期间,五名跑步者中的一名成功收集到了无噪声或伪迹的数据。在马拉松比赛的最初一小时内,其余四名跑步者因电极与皮肤的粘附性差而阻碍了数据收集,这导致与第一小时相比采集率显著下降(从86.7±13.4%降至37.3±36.9%,P = 0.028)。在马拉松比赛期间,一名跑步者检测到三导联中具有清晰ECG波形的室性早搏成对出现。需要对设备进行进一步改进,以使马拉松跑步者能够在不影响其表现的情况下高效获取ECG。本研究还展示了三导联可穿戴ECG监测在其他短时间运动和远程家庭心脏康复中的潜在应用。
这是一项在全程马拉松比赛期间使用三导联可穿戴心电图(ECG)监测设备的初步试验。在马拉松比赛的第一小时内以低噪声和伪迹获得了ECG数据;然而,由于电极粘附性差,在中后期数据采集率下降。本研究证明了在短期运动和心脏康复期间应用可穿戴ECG监测以检测预警信号和预防心脏骤停的可能性。