Suppr超能文献

一项关于瓦格纳表面置换关节成形术中植入物失败情况的研究。

A study of implant failure in the Wagner resurfacing arthroplasty.

作者信息

Bell R S, Schatzker J, Fornasier V L, Goodman S B

出版信息

J Bone Joint Surg Am. 1985 Oct;67(8):1165-75.

PMID:3902844
Abstract

Using clinical, radiographic, and pathological data, we investigated eighteen cases of early aseptic failure of an implant in patients who had undergone reconstruction of the hip with a Wagner resurfacing prosthesis. Sixteen patients required revision for loosening of the acetabular component, with eight of them also demonstrating loosening of the femoral component. One patient had loosening of the femoral component without failure of the acetabular component, and one patient sustained a femoral neck fracture that was associated with osteonecrosis. Six of the patients with loosening of the acetabular component had an associated significant loss of acetabular bone stock. Loosening was associated with the development of a membrane at the bone-cement interface in all patients. Histological examination of the membrane demonstrated a marked foreign-body response to wear products from the arthroplasty. Bone resorption appeared active at the bone-membrane interface. We concluded that the acetabular component of the Wagner prosthesis is prone to early loosening and that the early loosening is potentiated by a foreign-body response to debris resulting from arthroplastic wear.

摘要

利用临床、影像学和病理学数据,我们对18例接受Wagner表面置换假体髋关节重建术的患者早期植入物无菌性失败的病例进行了研究。16例患者因髋臼部件松动需要翻修,其中8例还伴有股骨部件松动。1例患者股骨部件松动但髋臼部件未失效,1例患者发生股骨颈骨折并伴有骨坏死。髋臼部件松动的患者中有6例伴有髋臼骨量显著丢失。所有患者的松动均与骨水泥界面处的一层膜的形成有关。对该膜的组织学检查显示对关节成形术磨损产物有明显的异物反应。骨吸收在骨 - 膜界面处表现活跃。我们得出结论,Wagner假体的髋臼部件易于早期松动,并且这种早期松动因对关节成形术磨损产生的碎屑的异物反应而加剧。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验