Narita Akihiro, Ohkubo Masaki, Ohsugi Yuki
Department of Radiological Technology, Graduate School of Health Sciences, Niigata University.
Division of Radiology, Niigata University Medical and Dental Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi. 2024 Sep 20;80(9):923-927. doi: 10.6009/jjrt.2024-1486. Epub 2024 Aug 15.
The measurement of slice sensitivity profile (SSP) in non-helical CT is conventionally performed by repeated scans with moving a micro-coin phantom little by little in the longitudinal direction at a small interval, which is reliable but laborious and time-consuming. The purpose of this study was to propose a simple method for measuring the SSP in non-helical CT based on a previous method that measured the slice thickness using a tilted metal wire.
In the proposed method, a CT image was obtained by scanning a wire tilted at an angle θ=30° to the scan plane. By deconvolving the image with the point spread function (PSF) measured at the scanner, we obtained an image that was not affected by the PSF blurring. The CT value profile along the wire was acquired on the obtained image. The SSP was determined by multiplying the profile by tan θ. In addition, the SSP was measured by the conventional method using a micro-coin phantom and compared with the SSP obtained by the proposed method.
The SSP measured by the proposed method agreed well with that measured by the conventional method. The full-width at half-maximum values of these SSPs were the same.
The proposed method was demonstrated to easily and accurately measure the SSP in non-helical CT.