Interdisciplinary Laboratory of Clinical Analysis, Interlab-UMU, Regional Campus of International Excellence Campus Mare Nostrum, University of Murcia, Murcia, 30100, Spain.
Department of Animal Production, Regional Campus of International Excellence 'Campus Mare Nostrum', University of Murcia, Campus de Espinardo S/N, Murcia, 30100, Spain.
BMC Vet Res. 2024 Aug 19;20(1):370. doi: 10.1186/s12917-024-04195-5.
Cortisone is derived from cortisol through the action of the enzyme 11β-hydroxysteroid dehydrogenase type II, and it has gained importance in recent years as a biomarker of stress. This study aimed to develop and validate an assay for the measurement of cortisone in pig saliva and evaluate whether its concentration varies in stressful situations. For this purpose, a specific immunoassay was developed and validated analytically, and a study was performed to evaluate whether cortisone concentrations in saliva can vary under heat stress conditions.
The assay proved to be accurate, reliable, and sensitive for the measurement of cortisone in pig saliva. The limit of detection of the assay was set at 0.006 ng/ml, and the lower limit of quantification was 0.023 ng/ml. It also correlated significantly with the results obtained by LC‒MS/MS (P = 0.003; r = 0.64). In addition, the cortisone concentration in animals subjected to prolonged heat stress decreased significantly 15 days after treatment (P < 0.0001).
According to these results, cortisone measured by this assay could be used as a tool for the non-invasive evaluation of thermal stress in pig saliva.
皮质酮通过酶 11β-羟甾类脱氢酶 II 的作用从皮质醇中衍生而来,近年来作为应激的生物标志物变得越来越重要。本研究旨在开发和验证一种用于测量猪唾液中皮质酮的方法,并评估其在应激情况下的浓度是否会发生变化。为此,开发并验证了一种特定的免疫测定法,并进行了一项研究以评估唾液中皮质酮的浓度是否会在热应激条件下发生变化。
该测定法被证明对猪唾液中皮质酮的测量具有准确性、可靠性和敏感性。该测定法的检测限设定为 0.006ng/ml,定量下限为 0.023ng/ml。它与 LC-MS/MS 获得的结果也显著相关(P=0.003;r=0.64)。此外,在治疗 15 天后,长期受热应激的动物的皮质酮浓度显著降低(P<0.0001)。
根据这些结果,通过该测定法测量的皮质酮可用于非侵入性评估猪唾液中的热应激。