Accuracy of an image matching technique for assessing knee alignment during the stance phase of gait using single-plane anteroposterior radiography.

作者信息

Kobayashi Koichi, Sakamoto Makoto, Soeno Tatsuya, Sato Takashi

机构信息

School of Health Sciences, Faculty of Medicine, Niigata University, Niigata, Japan.

Department of Orthopedic Surgery, Niigata Medical Center, Niigata, Japan.

出版信息

Biomed Mater Eng. 2024 Nov;35(6):522-529. doi: 10.3233/BME-240059.

Abstract

BACKGROUND

Accurate measurement of bone alignment of the knee during walking provides ideal clinical information for diagnosis and treatment of knee joint disorders. To bring this ideal closer to clinical reality, we developed an image matching technique to measure the three-dimensional (3D) position of bones using anteroposterior radiography during the stance phase of treadmill walking.

OBJECTIVE

To develop and validate an image matching method for evaluation of 3D knee alignment using anteroposterior radiography of artificial femoral and tibial bones.

METHODS

The 3D position of each bone was recovered by minimizing the difference between the projected outline and the contour of the bone in the anteroposterior radiograph. The true value of the position was measured using a 3D coordinate measuring machine.

RESULTS

The mean values ± standard deviation and root mean squares (RMS) of translation errors were within -1.6 ± 2.1 mm and 2.6 mm, respectively, for femur, and 2.1 ± 1.9 mm and 2.8 mm for tibia. The mean values ± standard deviation and RMS of errors in rotation were within 0.3 ± 0.7° and 0.7°, respectively, for femur, and -0.3 ± 0.9° and 0.9°, respectively, for tibia.

CONCLUSION

Our method is suitable for evaluating 3D knee alignment on anteroposterior radiography.