Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study Investigating Temperature Changes and Bonding Efficacy.

作者信息

Sfondrini Maria Francesca, Pascadopoli Maurizio, Gandini Paola, Preda Lorenzo, Sfondrini Domenico, Bertino Karin, Rizzi Cinzia, Scribante Andrea

机构信息

Unit of Orthodontics and Pediatric Dentistry, Section of Dentistry, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy.

Diagnostic Imaging and Radiotherapy Unit, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy.

出版信息

Dent J (Basel). 2024 Aug 13;12(8):255. doi: 10.3390/dj12080255.

Abstract

OBJECTIVES

Orthodontists are often asked to remove fixed retainers before patients undergo magnetic resonance imaging (MRI). The present in vitro study was designed to analyze the heating and bonding efficacy of stainless steel multibraided fixed retainers after 1.5- and 3-tesla (T) MRI.

MATERIALS AND METHODS

A total of 180 human mandibular incisors were used to create 45 specimens of four teeth each, divided into nine groups. Handmade multibraided fixed retainers of three different sizes, defined by the diameter of the initial wire used (0.008″, 0.010″ and 0.012″), were tested. Three groups underwent MRI at 1.5 T, another three groups underwent MRI at 3 T and the last three groups did not undergo MRI. Temperature was assessed before and after MRI. Shear bond strength (SBS) and adhesive remnant index (ARI) were assessed after MRI for all groups. Data were statistically analyzed ( < 0.05).

RESULTS

After 1.5 T exposure, no significant temperature increase from T0 to T1 was observed in any of the groups ( > 0.05). Regarding the 3 T groups, a significant difference from T0 to T1 was found for all the groups ( < 0.05). Temperature changes were not clinically relevant, as they were less than 1 °C for all groups except for group 3 (ΔT0-T1: 1.18 ± 0.3 °C) and group 6 (ΔT0-T1: 1.12 ± 0.37 °C). Furthermore, there were no significant differences between the temperature variations associated with different wire diameters ( > 0.05).

CONCLUSIONS

No significant changes in SBS or ARI were found ( > 0.05).

CLINICAL SIGNIFICANCE

Since overheating was irrelevant and adhesion values did not change, the tested devices were concluded to be safe for MRI examinations at 1.5 T and 3 T.

摘要

目的

正畸医生经常被要求在患者进行磁共振成像(MRI)之前拆除固定保持器。本体外研究旨在分析1.5特斯拉(T)和3T MRI后不锈钢多股编织固定保持器的加热情况和粘结效果。

材料与方法

共使用180颗人类下颌切牙制作45个标本,每个标本包含四颗牙齿,分为九组。测试了三种不同尺寸的手工多股编织固定保持器,其尺寸由初始钢丝的直径定义(0.008英寸、0.010英寸和0.012英寸)。三组在1.5T下进行MRI,另外三组在3T下进行MRI,最后三组未进行MRI。在MRI前后评估温度。对所有组在MRI后评估剪切粘结强度(SBS)和粘结残留指数(ARI)。对数据进行统计学分析(<0.05)。

结果

在1.5T暴露后,任何组从T0到T1均未观察到显著的温度升高(>0.05)。对于3T组,所有组从T0到T1均发现有显著差异(<0.05)。温度变化与临床无关,因为除了第3组(ΔT0-T1:1.18±0.3℃)和第6组(ΔT0-T1:1.12±0.37℃)外,所有组的温度变化均小于1℃。此外,不同钢丝直径相关的温度变化之间无显著差异(>0.05)。

结论

未发现SBS或ARI有显著变化(>0.05)。

临床意义

由于过热无关紧要且粘结值未改变,得出测试装置在1.5T和3T的MRI检查中是安全的结论。