Bhagat Esha, Agnihotri Yoshaskam, Das Abhisek, Das Sambarta, Panda Swagat, Hota Sudeepta
Conservative Dentistry and Endodontics, Hi-Tech Dental College and Hospital, Bhubaneswar, IND.
Cureus. 2024 May 14;16(5):e60283. doi: 10.7759/cureus.60283. eCollection 2024 May.
Traditionally, radiographs were used to assess the working length of root canals. However, the use of Electronic Apex Locators (EAL) may be a non-invasive alternative. This study aimed to evaluate the accuracy of the electronic apex locator (EAL) compared to radiographic measurements in determining the working length of root canals.
The study assessed the performance of EALs in different scenarios, including the presence of blood and pulp tissue, the use of ethylenediaminetetraacetic acid (EDTA), sodium hypochlorite (NaOCl) as an irrigant, and post-instrumentation with normal saline. An electronic apex locator (DTE DPEX-V; Woodpecker, China) was used alongside traditional radiographs to measure the working length in 144 root canal treatments. Bland-Altman analysis was used to compare the measurements between EAL and radiographs using Jamovi software, v2.4.8 (https://www.jamovi.org/).
The findings revealed strong correlations between EAL and radiographic measurements across all testing environments, with Pearson's correlation coefficients ranging from 0.944 to 0.998. The Bland-Altman analysis suggests bias; the average difference was negative and close to zero (Pulp: -0.354, EDTA: -0.0972, NaOCl: -0.0382, Saline: -0.0139) when EAL measurements were compared to radiographic measurements.
The working length measurement of root canals using an electronic apex locator such as DTE DPEX-V is comparable to traditional radiographic measurements. The use of irrigants improves the measurement. The EAL has the potential to be an alternative to the invasive radiograph for root canal length measurement.
传统上,X线片用于评估根管工作长度。然而,使用电子根尖定位仪(EAL)可能是一种非侵入性的替代方法。本研究旨在评估电子根尖定位仪(EAL)与X线片测量相比在确定根管工作长度方面的准确性。
本研究评估了EAL在不同情况下的性能,包括存在血液和牙髓组织、使用乙二胺四乙酸(EDTA)、次氯酸钠(NaOCl)作为冲洗剂以及用生理盐水进行根管预备后。使用电子根尖定位仪(DTE DPEX-V;啄木鸟,中国)与传统X线片一起测量144例根管治疗中的工作长度。使用Jamovi软件v2.4.8(https://www.jamovi.org/)进行Bland-Altman分析,以比较EAL和X线片之间的测量结果。
研究结果显示,在所有测试环境中,EAL与X线片测量之间存在强相关性,Pearson相关系数范围为0.944至0.998。Bland-Altman分析表明存在偏差;当将EAL测量结果与X线片测量结果进行比较时,平均差异为负且接近零(牙髓:-0.354,EDTA:-0.0972,NaOCl:-0.0382,生理盐水:-0.0139)。
使用DTE DPEX-V等电子根尖定位仪测量根管工作长度与传统X线片测量相当。冲洗剂的使用改善了测量结果。EAL有可能成为根管长度测量中侵入性X线片的替代方法。