Wu Yingyi, Ye Zheng, Yang Ting, Yao Shan, Chen Jie, Yin Ting, Tang Hehan, Song Bin
Department of Radiology, West China Hospital, Sichuan University, Chengdu, China.
MR Collaborations, Siemens Healthineers Ltd., Chengdu, China.
Quant Imaging Med Surg. 2024 Sep 1;14(9):6684-6697. doi: 10.21037/qims-24-341. Epub 2024 Aug 28.
Simultaneous multislice (SMS) technology improves acquisition efficiency of diffusion-weighted imaging (DWI). This study aimed to evaluate the performance of SMS-DWI in image quality and apparent diffusion coefficient (ADC) measurements for focal liver lesions (FLLs) as compared with that of conventional DWI (CON-DWI).
The institutional ethics committee of West China Hospital, Sichuan University approved this single-center, prospective study conducted from February 2021 to March 2022. Free-breathing SMS-DWI and CON-DWI examinations were acquired on a 3-T scanner with b-values of 50, 400, and 800 s/mm. Qualitative image quality and quantitative measurements of signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and ADC were compared between SMS-DWI and CON-DWI. The ADC values for FLLs were further compared between SMS-DWI and CON-DWI in different patient subgroups. The intra- and interreader agreements were assessed. Significance was set at P<0.05.
This study included 116 patients (96 males, 20 females; mean age 52.0±10.7 years) with 119 FLLs. No significant differences were observed between SMS-DWI and CON-DWI regarding overall image quality in any b-value DWIs, and there were also no differences observed between SMS-DWI and CON-DWI (b=800 s/mm) for either SNR or CNR (both P values >0.05). ADC values obtained from CON-DWI were higher than those from SMS-DWI in all FLLs [(1.31±0.47)×10 (1.26±0.46)×10 mm/s; P=0.004], and similar findings were observed across the different patient subgroups. The consistency analysis showed intrareader intraclass correlation coefficient (ICC) values of 0.792-0.944 and interreader ICC values of 0.758-0.861 for quantitative measurements (SNR, CNR, and ADC) and kappa values of 0.609-0.878 for qualitative image quality.
SMS-DWI achieved a 37% reduction in scan time compared to CON-DWI while maintaining comparable overall image quality. Notably, the ADC values for FLLs were observed to be quantitatively lower with SMS-DWI.
同时多层(SMS)技术提高了扩散加权成像(DWI)的采集效率。本研究旨在评估与传统DWI(CON-DWI)相比,SMS-DWI在局灶性肝病变(FLL)的图像质量和表观扩散系数(ADC)测量方面的性能。
四川大学华西医院伦理委员会批准了这项于2021年2月至2022年3月进行的单中心前瞻性研究。在一台3-T扫描仪上进行自由呼吸的SMS-DWI和CON-DWI检查,b值分别为50、400和800 s/mm²。比较了SMS-DWI和CON-DWI之间的定性图像质量以及信号噪声比(SNR)、对比噪声比(CNR)和ADC的定量测量结果。进一步比较了不同患者亚组中SMS-DWI和CON-DWI对FLL的ADC值。评估了阅片者内和阅片者间的一致性。显著性设定为P<0.05。
本研究纳入了116例患者(96例男性,20例女性;平均年龄52.0±10.7岁),共119个FLL。在任何b值的DWI中,SMS-DWI和CON-DWI在整体图像质量方面均未观察到显著差异,在SNR或CNR方面,SMS-DWI和CON-DWI(b = 800 s/mm²)之间也未观察到差异(P值均>0.05)。在所有FLL中,CON-DWI获得的ADC值高于SMS-DWI [(1.31±0.47)×10⁻³ vs. (1.26±0.46)×10⁻³ mm²/s;P = 0.004],在不同患者亚组中也观察到了类似的结果。一致性分析显示,定量测量(SNR、CNR和ADC)的阅片者内组内相关系数(ICC)值为0.792 - 0.944,阅片者间ICC值为0.758 - 0.861,定性图像质量的kappa值为0.609 - 0.878。
与CON-DWI相比,SMS-DWI扫描时间减少了37%,同时保持了相当的整体图像质量。值得注意的是,SMS-DWI测得的FLL的ADC值在定量上较低。