Département des Sciences Cliniques des Animaux de Compagnie et de Sport, Université de Toulouse, ENVT, Toulouse, France.
CREFRE, Université de Toulouse, INSERM, ENVT, UPS, Toulouse, France.
Vet Clin Pathol. 2024 Sep;53(3):294-308. doi: 10.1111/vcp.13377. Epub 2024 Sep 18.
The Sysmex XN-V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT-2000iV analyzer include an optic-fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count.
We aimed to validate the Sysmex XN-V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations.
Feline EDTA blood specimens were analyzed on the Sysmex XN-V to evaluate repeatability, linearity, comparison with the XT-2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT-2000iV RIs.
Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN-V and XT-2000iV analyzers was good (≥0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN-V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT-2000iV RIs can be used to interpret results from the Sysmex XN-V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN-V measurements after storage at 4 and 24°C were similar to those described for the Sysmex XT-2000iV analyzer.
The performance of the Sysmex XN-V analyzer was good and compared favorably with the Sysmex XT-2000iV analyzer.
XN-V 源自新型 Sysmex XN 系列人类血液分析仪。与之前经过验证的 XT-2000iV 分析仪相比,主要变化包括对血小板进行光学荧光分析和有核红细胞 (NRBC) 计数。
根据美国兽医临床病理学会和国际血液学标准化委员会的建议,我们旨在对猫的血液进行 Sysmex XN-V 的验证。
对 Sysmex XN-V 上的猫 EDTA 血液样本进行分析,以评估重复性、线性、与 XT-2000iV 分析仪和手动方法的比较、稳定性,并验证之前建立的 Sysmex XT-2000iV RI。
大多数变量的重复性均非常好。除嗜酸性粒细胞和血小板变量外,大多数变量的目测线性均非常好或良好。XN-V 与 XT-2000iV 分析仪之间的相关性非常好(≥0.82),除网织红细胞指数外,所有变量均如此。XN-V 与手动差异计数之间的相关性大多为良好到极好,中性粒细胞为可接受,单核细胞和 NRBC 为一般。除红细胞分布宽度和网织红细胞变量外,大多数变量都可以使用之前建立的 Sysmex XT-2000iV RI 来解释 XN-V 分析仪的结果。由于血小板聚集体,无法评估血小板变量的 RI。XN-V 测量值在 4°C 和 24°C 储存后的变化与 Sysmex XT-2000iV 分析仪的描述相似。
XN-V 分析仪的性能良好,与 Sysmex XT-2000iV 分析仪相比表现出色。