Hill S F, Haldane D J, Ngui-Yen J H, Smith J A
J Clin Microbiol. 1985 Sep;22(3):465-6. doi: 10.1128/jcm.22.3.465-466.1985.
We compared susceptibility tests of 47 Pseudomonas aeruginosa isolates and 40 Pseudomonas species to carbenicillin and trimethoprim-sulfamethoxazole by the MS-2 and Sceptor systems and agar dilution. The major and very major errors encountered in these tests in the MS-2 and Sceptor systems raise doubts about the accuracy of these methods for testing P. aeruginosa and confirm that they should not be used for testing the susceptibility of Pseudomonas species to the two drugs tested.
我们通过MS-2和Sceptor系统以及琼脂稀释法,比较了47株铜绿假单胞菌分离株和40株假单胞菌对羧苄青霉素和甲氧苄啶-磺胺甲恶唑的药敏试验。在MS-2和Sceptor系统的这些试验中遇到的主要和非常主要的误差,让人对这些方法检测铜绿假单胞菌的准确性产生怀疑,并证实它们不应用于检测假单胞菌对所测试的两种药物的敏感性。