Transcutaneous oxygen and carbon dioxide monitoring in sick neonates using a combined sensor.

作者信息

Bhat R, Diaz-Blanco J, Chaudhry U, Vidyasagar D

出版信息

Chest. 1985 Dec;88(6):890-4. doi: 10.1378/chest.88.6.890.

Abstract

Using a combined O2-CO2 sensor, we evaluated the effect of cathode size and membrane thickness and incorporation of correction factor upon transcutaneous O2-CO2 readings. The present studies were carried out in three phases in newborns less than a week of age. The data from these studies indicate that a combined O2-CO2 sensor with a smaller cathode, Teflon membrane and with a built-in correction factor can accurately reflect arterial O2 and CO2 tension and can replace the present two heated sensors.