Kendrisic Milana, Nienhaus Jonas, Agafonov Vladislav, Salas Matthias, Nguyen Quang, Resch Hemma, Vass Clemens, Drexler Wolfgang, Schmoll Tilman, Leitgeb Rainer A
Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna 1090, Austria.
Christian Doppler Laboratory for Innovative Optical Imaging and its Translation to Medicine, Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna 1090, Austria.
Biomed Opt Express. 2024 Jul 17;15(8):4719-4736. doi: 10.1364/BOE.523081. eCollection 2024 Aug 1.
Low-cost optical coherence tomography has recently emerged as a growing field due to the increased need for general availability of OCT devices outside of the clinics. One of the main obstacles in creating low-cost SS-OCT systems is the price of the laser. In this work, we study the influence of different tuning parameters (e.g., frequency, duty cycle, modulation curve, temperature) on the resulting bandwidth of the previously proposed low-cost single-mode thermally-tunable vertical-cavity surface-emitting laser (VCSEL) source at 850 nm. With optimal parameters, the laser achieves a tuning bandwidth of 10.2 nm at a 50 kHz A-scan rate. In addition, we show the first 3D rendered volume scans of both anterior and posterior segment using a custom VCSEL-based low-cost OCT setup. With the help of deep-learning-based denoising, it was possible to critically reduce the noise in single scans. Moreover, by investigating the phase stability, it became apparent that phase stability between sweeps increases with rising modulation frequencies, making the auxiliary interferometer obsolete. Thus, the system's 50 kHz tuning regimen is also suitable for functional extensions such as OCT angiography.
由于临床外对光学相干断层扫描(OCT)设备普遍可用性的需求增加,低成本光学相干断层扫描最近已成为一个不断发展的领域。制造低成本扫频光学相干断层扫描(SS-OCT)系统的主要障碍之一是激光器的价格。在这项工作中,我们研究了不同调谐参数(例如频率、占空比、调制曲线、温度)对先前提出的850纳米低成本单模热调谐垂直腔面发射激光器(VCSEL)源的所得带宽的影响。通过优化参数,该激光器在50千赫兹A扫描速率下实现了10.2纳米的调谐带宽。此外,我们展示了使用基于定制VCSEL的低成本OCT装置对前段和后段进行的首次三维渲染容积扫描。借助基于深度学习的去噪技术,可以显著降低单次扫描中的噪声。此外,通过研究相位稳定性,很明显扫描之间的相位稳定性随着调制频率的升高而增加,使得辅助干涉仪不再必要。因此,该系统的50千赫兹调谐方案也适用于诸如OCT血管造影等功能扩展。