Eiermann Jennifer S, Hazuchova Katarina, Vidondo Beatriz, Campos Miguel, Schuller Simone, Cui Yi
Division of Small Animal Internal Medicine, Department of Clinical Veterinary Science, Vetsuisse Faculty, University of Bern, 3012 Bern, Switzerland.
Department of Veterinary Clinical Sciences, Small Animal Clinic, Internal Medicine, Justus-Liebig University Giessen, 35392 Giessen, Germany.
Animals (Basel). 2024 Sep 26;14(19):2787. doi: 10.3390/ani14192787.
Cats with diabetic ketosis or ketoacidosis DK(A) require intensive glucose monitoring. The aim of this study was to assess the agreement between a portable blood glucose meter (PBGM) and a flash glucose monitoring system (FGMS; FreeStyle Libre 2.0 Abbott) measuring interstitial glucose in cats with DK(A). Ten client-owned cats with naturally occurring DK(A) were prospectively enrolled. Glucose concentrations were assessed with both methods every 1-3 h until resolution of DK(A), and every 4-8 h thereafter. While the median FGMS measured glucose concentration (14.3 mmol/L) was significantly lower than the median PBGM measured glucose concentration (19 mmol/L) ( < 0.001), the overall correlation between the FGMS and PBGM was high (r = 0.88; < 0.001). In the Parkes error grid analysis, 98.3% of measurements fell in zones A and B. Bland-Altman plot analysis demonstrated that in the low glycaemic range (BG < 5.5 mmol/L), 50% of FGMS measurements deviated more than ±0.83 mmol/L, and in the high glycaemic range (BG > 5.5 mmol/L), 81% of results deviated >15% from the PBGM measurements. There was significant inter-individual variation in the difference between glucose concentrations measured by the FGMS and PBGM ( < 0.001). In spite of being more easily tolerated and easier to use, currently this method cannot be recommended for routine monitoring of cats with DK(A).
患有糖尿病酮症或酮症酸中毒(DK(A))的猫需要进行强化血糖监测。本研究的目的是评估便携式血糖仪(PBGM)与测量患有DK(A)的猫的组织间液葡萄糖的闪光葡萄糖监测系统(FGMS;雅培Freestyle Libre 2.0)之间的一致性。前瞻性纳入了10只患有自然发生的DK(A)的客户拥有的猫。在DK(A)消退之前,每1 - 3小时用两种方法评估一次血糖浓度,之后每4 - 8小时评估一次。虽然FGMS测量的血糖浓度中位数(14.3 mmol/L)显著低于PBGM测量的血糖浓度中位数(19 mmol/L)(<0.001),但FGMS与PBGM之间的总体相关性较高(r = 0.88;<0.001)。在帕克斯误差网格分析中,98.3%的测量值落在A区和B区。布兰德 - 奥特曼图分析表明,在低血糖范围(血糖<5.5 mmol/L),50%的FGMS测量值偏差超过±0.83 mmol/L,在高血糖范围(血糖>5.5 mmol/L),81%的结果与PBGM测量值偏差>15%。FGMS和PBGM测量的葡萄糖浓度差异存在显著的个体间差异(<0.001)。尽管该方法更容易耐受且使用更方便,但目前不推荐将其用于患有DK(A)的猫的常规监测。