不同光照条件对使用口腔内扫描仪进行牙齿比色的影响:一项研究。

Impact of different lighting conditions on the tooth-shade selection using intra-oral scanners: An study.

作者信息

Nantanapiboon Dusit, Kamnoedboon Porawit, Srinivasan Murali

机构信息

Clinic of General-, Special Care-, and Geriatric Dentistry, Center for Dental Medicine, University of Zurich, Zurich, Switzerland.

Department of Operative Dentistry, Faculty of Dentistry, Chulalongkorn University, Bangkok, 10330, Thailand.

出版信息

Heliyon. 2024 Oct 2;10(19):e38870. doi: 10.1016/j.heliyon.2024.e38870. eCollection 2024 Oct 15.

Abstract

OBJECTIVES

To evaluate the accuracy in tooth shade selection under various lighting conditions using different devices, including a color spectrophotometer (CSP) and two intra-oral scanners (IOS: IOS-1 and IOS-2).

METHODS

Tooth shade measurements were performed on a standardized shade guide (16 shades, A1-D4) using CSP, IOS-1 and IOS-2. These evaluations were carried out under three lighting conditions: (1) DL - device light source only, (2) RL - DL plus room light, and (3) CL - RL plus dental-chair light. Each shade was measured ten times per lighting condition with each device by a single investigator. Shade detection accuracy was defined as the percentage of matches to the known shade. Logistic regression models were used for statistical analyses ( = 0.05).

RESULTS

A total of 1440 measurements were conducted. Lighting conditions of increasing intensity significantly decreased the shade detection accuracy of the devices (CSP: RLvsDL, p > 0.9999, Odds Ratio (OR) = 1; CLvsDL, p = 0.0726, OR = 0.6053, IOS-1: RLvsDL, p < 0.0001, OR = 0.2862; CLvsDL, p < 0.0001, OR = 0.05239 and IOS-2: RLvsDL, p = 0.0002, OR = 0.376; CLvsDL, p < 0.0001, OR = 0.1908). CSP, IOS-1, and IOS-2 had shade detection accuracies of 43.75 %, 74.38 %, and 59.38 % respectively under DL. In RL, IOS-1 and IOS-2 dropped to 50.63 % and 40.63 %, respectively; CSP remained unchanged. Under CL, the accuracy dropped for all devices (CSP: 35.63 %, IOS-1: 20.63 %, IOS-2: 28.75 %). Additionally, the devices demonstrated higher accuracy when measuring shade A compared to shades B, C, and D.

CONCLUSIONS

The findings of this in vitro study concluded that the accuracy of tooth shade detection was more reliable, particularly for individuals with shade A, when spectrophotometers and intraoral scanners were used with only their built-in light source. Further purpose-built clinical studies are required to confirm these findings in a clinical context.

摘要

目的

使用不同设备(包括一台色彩分光光度计(CSP)和两台口腔内扫描仪(IOS:IOS - 1和IOS - 2))评估在各种光照条件下牙齿比色的准确性。

方法

使用CSP、IOS - 1和IOS - 2在标准化比色板(16种色阶,A1 - D4)上进行牙齿比色测量。这些评估在三种光照条件下进行:(1)DL - 仅设备光源,(2)RL - DL加室内光,(3)CL - RL加牙科椅灯光。每种设备在每种光照条件下,由一名研究人员对每个色阶测量10次。比色检测准确性定义为与已知色阶匹配的百分比。使用逻辑回归模型进行统计分析(α = 0.05)。

结果

共进行了1440次测量。光照强度增加的条件显著降低了设备的比色检测准确性(CSP:RL与DL相比,p > 0.9999,优势比(OR) = 1;CL与DL相比,p = 0.0726,OR = 0.6053,IOS - 1:RL与DL相比,p < 0.0001,OR = 0.2862;CL与DL相比,p < 0.0001,OR = 0.05239,IOS - 2:RL与DL相比,p = 0.0002,OR = 0.376;CL与DL相比,p < 0.0001,OR = 0.1908)。在DL条件下,CSP、IOS - 1和IOS - 2的比色检测准确性分别为43.75%、74.38%和59.38%。在RL条件下,IOS - 1和IOS - 2分别降至50.63%和40.63%;CSP保持不变。在CL条件下,所有设备的准确性均下降(CSP:35.63%,IOS - 1:20.63%,IOS - 2:28.75%)。此外,与测量色阶B、C和D相比,设备在测量色阶A时表现出更高的准确性。

结论

这项体外研究的结果得出结论,当分光光度计和口腔内扫描仪仅使用其内置光源时,牙齿比色检测的准确性更可靠,特别是对于色阶A的个体。需要进一步专门的临床研究在临床环境中证实这些发现。