皮褶厚度和运动强度对股外侧肌近红外光谱法可靠性的影响。

The impact of skinfold thickness and exercise intensity on the reliability of NIRS in the vastus lateralis.

作者信息

Stuer Lena, Teso Massimo, Colosio Alessandro L, Loi Maura, Mucci Patrick, Pogliaghi Silvia, Boone Jan, Caen Kevin

机构信息

Department of Movement and Sport Sciences, Ghent University, Watersportlaan 2, 9000, Ghent, Belgium.

Univ. Lille, Univ. Artois, Univ. Littoral Côte d'Opale, ULR, URePSSS-Unité de Recherche Pluridisciplinaire Sport Santé Société, 7369, 59000, Lille, France.

出版信息

Eur J Appl Physiol. 2025 Apr;125(4):1061-1073. doi: 10.1007/s00421-024-05654-5. Epub 2024 Nov 21.

Abstract

PURPOSE

The aims of this study were (1) to assess the test-retest reliability of the primary near infrared spectroscopy (NIRS) variables (i.e., StO, T[Hb], [HbO] and [HHb]) during cycling and (2) to investigate potential influences of exercise intensity and adipose tissue thickness (ATT) on this reliability.

METHODS

21 men and 20 women completed twelve constant work rate tests (6 min) at six different exercise intensities with each intensity performed twice. NIRS variables were measured at the vastus lateralis. The coefficient of variance (CV%), the intraclass correlation coefficient (ICC), mean bias and limits of agreement (LoA) were determined for reliability purposes.

RESULTS

The reliability of baseline values were acceptable to very good (CV% range: 5.83 - 21.96%). The reliability of end-values (CV% range: 0.02 - 25.02%, ICC range: 0.0 - 0.935) and amplitudes (CV% range: 0.46 - 5099%, ICC range: 0.0 - 0.887) were more variable. In general, the mean biases of end-values and amplitudes showed wide limits of agreement. A homogeneous influence of exercise intensity on reliability could not be established but reliability measures appeared to be lower in people with a lower skinfold thickness. Moreover, the NIRS signals decreased with increasing ATT but stabilized upon reaching a cut-off of 8 mm ATT. In addition, ATT did have a significant influence on [HHb] amplitude. In participants with ATT < 8 mm, higher amplitudes were observed with increasing intensity whereas in participants with ATT > 8 mm, there were no differences between the intensities.

CONCLUSION

The study reveals variable results with regards to reliability and there was no consistent influence of exercise intensity on reliability. Participants with a lower skinfold thickness showed stronger reliability. Moreover, NIRS signals decrease when ATT exceeds 8 mm. Careful consideration is necessary when interpreting NIRS signals in such cases.

摘要

目的

本研究的目的是:(1)评估骑行过程中主要近红外光谱(NIRS)变量(即StO、T[Hb]、[HbO]和[HHb])的重测信度;(2)研究运动强度和脂肪组织厚度(ATT)对该信度的潜在影响。

方法

21名男性和20名女性在六种不同运动强度下完成了十二次恒定工作率测试(6分钟),每种强度进行两次。在股外侧肌测量NIRS变量。为评估信度,确定了变异系数(CV%)、组内相关系数(ICC)、平均偏差和一致性界限(LoA)。

结果

基线值的信度为可接受至非常好(CV%范围:5.83 - 21.96%)。终末值(CV%范围:0.02 - 25.02%,ICC范围:0.0 - 0.935)和幅度(CV%范围:0.46 - 5099%,ICC范围:0.0 - 0.887)的信度变化更大。总体而言,终末值和幅度的平均偏差显示出较宽的一致性界限。运动强度对信度的均匀影响无法确定,但在皮褶厚度较低的人群中,信度测量值似乎较低。此外,NIRS信号随ATT增加而降低,但在ATT达到8毫米的临界值时趋于稳定。此外,ATT对[HHb]幅度有显著影响。在ATT<8毫米的参与者中,随着强度增加观察到更高的幅度,而在ATT>8毫米的参与者中,各强度之间没有差异。

结论

该研究揭示了关于信度的可变结果,且运动强度对信度没有一致的影响。皮褶厚度较低的参与者显示出更强的信度。此外,当ATT超过8毫米时,NIRS信号会降低。在这种情况下解释NIRS信号时需要仔细考虑。