Suppr超能文献

安全警示和注册异常数据衡量全膝关节植入物的安全性和性能的不同方面:安全警示和注册异常的比较研究。

Safety notices and registry outlier data measure different aspects of safety and performance of total knee implants: a comparative study of safety notices and register outliers.

机构信息

Department of Orthopaedics, Leiden University Medical Center, Leiden, the Netherlands.

Department of Electronics, Information and Biomedical Engineering, Politecnico di Milano, Milan, Italy.

出版信息

Acta Orthop. 2024 Nov 25;95:667-676. doi: 10.2340/17453674.2024.42361.

Abstract

BACKGROUND AND PURPOSE

Safety notices for medical devices such as total knee arthroplasty (TKA) implants may indicate problems in their design or performance that require corrective action to prevent patient harm. Safety notices are often published on national Ministries of Health or regulatory agencies websites. It is unknown whether problems triggering safety notices identify the same implants as those identified by registries as "outlier." We aimed to assess the extent to which safety notices and outlier identification in registries signal the same or different TKA implants.

METHODS

The CORE-MD tool, an automated web scraper tool, was used to collect safety notices related to TKA implants on 13 national Ministries of Health websites and regulatory agencies. Safety notices were defined according to the Medical Device Regulation (MDR) as "a communication sent by a manufacturer to users or customers in relation to a field safety corrective action." Identified TKA outliers, defined as having a significantly higher revision risk than other comparable TKA implants, were extracted from registry reports.

RESULTS

787 safety notices for 38 TKA implants and 35 TKA outliers were identified, together identifying 47 unique TKA implants. 26 (55%) TKA implants had safety notices and were also outliers, 12 (26%) TKA implants had only safety notices, and 9 (19%) were outliers only. TKA implants with safety notices only had similar types of problems to TKA outliers with safety notices, with "Manufacturing/Packaging/Shipping" problems being most frequent (44%). Cumulative revision risks (1/5/10 years) were lower for TKA implants with safety notices only than for TKA outliers with safety notices.

CONCLUSION

55% of the TKA with a safety notice were identified as outliers in the registry, whereas around 25% of TKA outliers are not the subject of publicly released safety notices, with safety notices pointing to TKA implants not identified by registries as potentially having a higher risk of failure. This suggests that safety notices and registry outlier data measure different aspects of safety and performance.

摘要

背景与目的

诸如全膝关节置换术(TKA)植入物等医疗器械的安全通知可能表明其设计或性能存在问题,需要采取纠正措施以防止患者受到伤害。安全通知通常在国家卫生部或监管机构的网站上发布。尚不清楚引发安全通知的问题是否与登记处确定为“异常值”的相同植入物相同。我们旨在评估安全通知和登记处确定的异常值在多大程度上识别出相同或不同的 TKA 植入物。

方法

使用 CORE-MD 工具(一种自动化网络抓取工具)从 13 个国家卫生部网站和监管机构收集与 TKA 植入物相关的安全通知。安全通知是根据医疗器械法规(MDR)定义的,即“制造商向用户或客户发送的与现场安全纠正措施有关的通信”。从登记报告中提取被确定为具有明显更高翻修风险的 TKA 异常值。

结果

共确定了 38 个 TKA 植入物和 35 个 TKA 异常值的 787 条安全通知,共确定了 47 个独特的 TKA 植入物。26(55%)个 TKA 植入物既有安全通知,又是异常值,12(26%)个 TKA 植入物只有安全通知,9(19%)个只有异常值。只有安全通知的 TKA 植入物与有安全通知的 TKA 异常值具有相似类型的问题,其中“制造/包装/运输”问题最为常见(44%)。只有安全通知的 TKA 植入物的累积翻修风险(1/5/10 年)低于有安全通知的 TKA 异常值。

结论

登记处中,55%带有安全通知的 TKA 被确定为异常值,而大约 25%的 TKA 异常值未发布公开安全通知,安全通知指出了未被登记处确定为潜在失败风险更高的 TKA 植入物。这表明安全通知和登记处异常值数据测量了安全性和性能的不同方面。