Jhuma Farjana Akter, Harada Kentaro, Misran Muhamad Affiq Bin, Mo Hin-Wai, Fujimoto Hiroshi, Hattori Reiji
Major of Device Science and Engineering, Interdisciplinary Graduate School of Engineering Sciences, Kyushu University, Fukuoka 8160811, Japan.
OPERA Solutions Inc., 5-5 Kyudai-Shimmachi, Nishi-ku, Fukuoka 8190388, Japan.
Sensors (Basel). 2024 Nov 29;24(23):7634. doi: 10.3390/s24237634.
A photoplethysmography (PPG) sensor is a cost-effective and efficacious way of measuring health conditions such as heart rate, oxygen saturation, and respiration rate. In this work, we present a hybrid PPG sensor system working in a reflective mode with an optoelectronic module, i.e., the combination of an inorganic light-emitting diode (LED) and a circular-shaped organic photodetector (OPD) surrounding the LED for efficient light harvest followed by the proper driving circuit for accurate PPG signal acquisition. The performance of the hybrid sensor system was confirmed by the heart rate detection process from the PPG using fast Fourier transform analysis. The PPG signal obtained with a 50% LED duty cycle and 250 Hz sampling rate resulted in accurate heart rate monitoring with an acceptable range of error. The effects of the LED duty cycle and the LED luminous intensity were found to be crucial to the heart rate accuracy and to the power consumption, i.e., indispensable factors for the hybrid sensor.
光电容积脉搏波描记法(PPG)传感器是一种经济高效的测量心率、血氧饱和度和呼吸频率等健康状况的方法。在这项工作中,我们展示了一种混合PPG传感器系统,它采用反射模式工作,配备一个光电模块,即一个无机发光二极管(LED)与围绕该LED的圆形有机光电探测器(OPD)相结合,以实现高效的光采集,随后是用于精确采集PPG信号的合适驱动电路。通过使用快速傅里叶变换分析从PPG进行心率检测过程,证实了混合传感器系统的性能。在50%的LED占空比和250 Hz采样率下获得的PPG信号实现了准确的心率监测,误差范围可接受。发现LED占空比和LED发光强度对心率准确性和功耗至关重要,即它们是混合传感器不可或缺的因素。