Maier E A, Rastegar F, Heimburger R, Ruch C, Pelletier A, Maier A, Leroy M J
Clin Chem. 1985 Apr;31(4):551-5.
We have applied energy dispersive x-ray fluorescence to the analysis of broncho-alveolar lavage fluids, to determine the concentration of several elements simultaneously with manipulation of the sample minimized. A 20-microL sample of a homogenized lavage fluid, to which two internal standards (containing Y and V) have been added, is deposited on a polypropylene film, 4 microns in thickness, and analyzed by x-ray fluorescence. We report the analytical procedure, the detection limits, and the reproducibility. The normal concentration range for trace elements is reported for a selected population. The simplicity and rapidity of the method, and the possibility of automating measurements, make this procedure suitable for screening large numbers of lavage specimens and to establish the diagnosis of some pneumoconioses.