Immanuel Jincy, Kongbrailatpam Tobias, Rajagopal Rohit, Simmons David
School of Medicine, Western Sydney University, Campbelltown, New South Wales, Australia.
College of Nursing, Texas Woman's University, Denton, Texas, USA.
Diabetes Obes Metab. 2025 Apr;27(4):2131-2137. doi: 10.1111/dom.16209. Epub 2025 Jan 24.
Physiological changes during pregnancy can infuence the performance of blood glucose meters. This study aimed to evaluate the analytical and clinical accuracy of glucose meters in pregnant women with hyperglycaemia.
Glucose was measured by four commonly used meters among consecutive women with diabetes in pregnancy. Capillary and venous samples were collected concurrently and compared with i-STAT (amperometry) and laboratory (hexokinase) glucose as reference methods. Bland-Altman plot, International Organization for Standardization criteria, surveillance error grid (SEG) and haematocrit influence were assessed.
In total, 824 paired samples from 103 women were analysed (GDM 57%, mean i-STAT capillary glucose 6.7 ± 2.3 mmol/L [121 ± 41 mg/dL], laboratory glucose 6.6 ± 2.4 mmol/L [119 ± 43 mg/dL], median haematocrit 0.36 L/L). Mean capillary glucose measured on all meters was significantly different from that measured on i-STAT (all p < 0.001), whereas venous glucose measured on Contour Next, Accu-Chek Guide and the laboratory (plasma) was similar. Contour Next had the lowest bias when using both reference methods (mean bias [95% limits of agreement] meter vs. i-STAT: Contour Next 1.3% [-8% to 10.6%], Accu-Chek Guide -3.2% [-11.4% to 5%], FreeStyle Optium Neo -11.9% [-24.7% to 0.8%] and LifeSmart 6.8% [-5.8% to 19.4%]; meter versus laboratory: -0.2% [-8.1% to 7.7%], -0.2% [-10.2% to 9.8%], -3.8% [-17.6% to 10%] and 6.1% [-5.9% to 18.2%]), respectively. Only Contour Next and Accu-Chek Guide had ≥97% of pairs within the SEG no-risk zone during both comparisons. Meters did not show haematocrit-related bias.
Accuracy of meters was higher when using venous samples than when using capillary samples. Contour Next and Accu-Chek Guide meters met accuracy standards in all analyses.
孕期的生理变化会影响血糖仪的性能。本研究旨在评估血糖仪在高血糖孕妇中的分析准确性和临床准确性。
在连续的妊娠糖尿病女性中,使用四种常用血糖仪测量血糖。同时采集毛细血管血样和静脉血样,并与作为参考方法的i-STAT(安培法)和实验室(己糖激酶法)血糖测量结果进行比较。评估了布兰德-奥特曼图、国际标准化组织标准、监测误差网格(SEG)和血细胞比容的影响。
共分析了103名女性的824对样本(妊娠期糖尿病占57%,i-STAT毛细血管血糖平均值为6.7±2.3 mmol/L [121±41 mg/dL],实验室血糖为6.6±2.4 mmol/L [119±43 mg/dL],血细胞比容中位数为0.36 L/L)。所有血糖仪测量的毛细血管血糖平均值与i-STAT测量值均有显著差异(所有p<0.001),而Contour Next、Accu-Chek Guide血糖仪和实验室(血浆)测量的静脉血糖相似。使用两种参考方法时,Contour Next的偏差最低(平均偏差[95%一致性界限]:血糖仪与i-STAT比较,Contour Next为1.3% [-8%至10.6%],Accu-Chek Guide为-3.2% [-11.4%至5%],FreeStyle Optium Neo为-11.9% [-24.7%至0.8%],LifeSmart为6.8% [-5.8%至19.4%];血糖仪与实验室比较,分别为-0.2% [-8.1%至7.7%],-0.2% [-10.2%至9.8%],-3.8% [-17.6%至10%],6.1% [-5.9%至18.2%])。在两次比较中,只有Contour Next和Accu-Chek Guide血糖仪有≥97%的配对样本处于SEG无风险区域。血糖仪未显示与血细胞比容相关的偏差。
使用静脉血样时血糖仪的准确性高于使用毛细血管血样时。Contour Next和Accu-Chek Guide血糖仪在所有分析中均符合准确性标准。