Biometry combining Scheimpflug imaging and partial coherence interferometry versus standard partial coherence interferometry biometry.

作者信息

van Renterghem V E W C M, Crahay F-X

机构信息

Centre Hospitalier Régionale Citadelle, Liège, Belgium.

Centre Hospitalier Régionale Citadelle, Liège, Belgium.

出版信息

J Fr Ophtalmol. 2025 Apr;48(4):104465. doi: 10.1016/j.jfo.2025.104465. Epub 2025 Mar 7.

Abstract

PURPOSE

To compare measurements of mean keratometry (K), axial length (AL), Jackson first and second astigmatism vector (J0 and J45), astigmatism magnitude, white-to-white (WTW) diameter and anterior chamber depth (ACD) values between two biometers, the Pentacam® AXL and the IOL Master 500.

SETTING

Public hospital in Belgium.

DESIGN

Retrospective instrument reliability analysis.

METHODS

Eyes of cataract or refractive lens exchange patients were examined preoperatively on a device combining Scheimpflug imaging and partial coherence interferometry (Pentacam® AXL) and a reference optical biometer (IOL Master 500). K, AL, J0, J45, astigmatism, WTW and ACD values were compared.

RESULTS

Three hundred and twenty-eight eyes of 187 cataract or refractive lens exchange patients (mean age 68years) were examined preoperatively. Measurements of mean keratometry, axial length, ACD, white-to-white and astigmatism were significantly different (P<0.05) between the two devices. Measurements of J0 and J45 did not significantly differ. Keratometry measurements with the Pentacam® AXL were significantly flatter than those measured with the IOL Master 500. Axial length measured with Pentacam® AXL was significantly greater when measured with the IOL Master 500, but this difference was not clinically relevant (0.015mm).

CONCLUSIONS

The Pentacam® AXL and the IOL Master 500 cannot be used interchangeably.

摘要

目的

比较两款生物测量仪(Pentacam® AXL和IOL Master 500)在平均角膜曲率(K)、眼轴长度(AL)、杰克逊第一和第二散光矢量(J0和J45)、散光量、白对白(WTW)直径和前房深度(ACD)值测量结果上的差异。

地点

比利时的公立医院。

设计

回顾性仪器可靠性分析。

方法

对白内障或屈光性晶状体置换患者的眼睛在术前使用一台结合了Scheimpflug成像和部分相干干涉测量技术的设备(Pentacam® AXL)以及一台参考光学生物测量仪(IOL Master 500)进行检查。比较K、AL、J0、J45、散光、WTW和ACD值。

结果

对187例白内障或屈光性晶状体置换患者(平均年龄68岁)的328只眼睛进行了术前检查。两款设备在平均角膜曲率、眼轴长度、ACD、白对白和散光的测量结果上存在显著差异(P<0.05)。J0和J45的测量结果无显著差异。Pentacam® AXL测量的角膜曲率明显比IOL Master 500测量的更平坦。用Pentacam® AXL测量的眼轴长度在用IOL Master 500测量时明显更长,但这种差异在临床上并不显著(0.015mm)。

结论

Pentacam® AXL和IOL Master 500不能互换使用。