Department of Ophthalmology and Visual Sciences (T.C.Y.C., K.H.W., F.Y.T., Y.M.W., and C.C.), The Chinese University of Hong Kong, Kowloon, Hong Kong; Hong Kong Eye Hospital (T.C.Y.C.), Kowloon, Hong Kong; and Department of Mathematics and Statistics (M.Y.), Hang Seng Management College, New Territories, Hong Kong.
Eye Contact Lens. 2020 Jan;46(1):35-45. doi: 10.1097/ICL.0000000000000603.
To compare the repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in cataract patients.
Three consecutive measurements were obtained using a swept-source biometer (IOLMaster 700) and a Scheimpflug biometer (AL-Scan) in 52 eyes of 52 patients. Keratometry, central corneal thickness (CCT), anterior chamber depth (ACD), axial length, and white-to-white (WTW) distance were recorded. Astigmatism values were transformed into vector components of J0 and J45. Intraoperator repeatability was analyzed using intraclass correlation coefficients (ICCs) and reproducibility coefficients (RCs). Agreement of measurements between the two devices was evaluated using the Bland-Altman method.
The IOLMaster 700 showed higher ICCs and lower RCs for the mean keratometry (Km) (P≤0.018), CCT (P≤0.027), and ACD (P≤0.001) measurements, whereas the AL-Scan showed higher ICC and lower RC for the J45 vector component of astigmatism at the 2.4-mm zone (P≤0.034). Both the devices had excellent repeatability (ICC=0.999) in axial length measurement. Systematic differences were found in Km, CCT, ACD, and WTW (P≤0.018) between the devices. The mean difference for Km was -0.196 and -0.144 D measured at the 2.4-mm zone and 3.3-mm zone, respectively. The corresponding mean difference for CCT, ACD, and WTW distance was 14.92 μm, -0.017 mm, and 0.283 mm, respectively. These differences led to a statistically significant but clinically insignificant difference in the prediction of intraocular lens power.
This study showed significant differences in anterior segment measurement repeatability and agreement between a swept-source biometer and a Scheimpflug biometer in eyes with cataract.
比较白内障患者中扫频源生物测量仪和 Scheimpflug 生物测量仪的重复性和一致性。
对 52 例(52 只眼)白内障患者连续 3 次使用扫频源生物测量仪(IOLMaster 700)和 Scheimpflug 生物测量仪(AL-Scan)进行测量。记录角膜曲率、中央角膜厚度(CCT)、前房深度(ACD)、眼轴长度和白到白(WTW)距离。散光值转换为 J0 和 J45 矢量分量。采用组内相关系数(ICC)和再现性系数(RC)分析操作者内重复性。采用 Bland-Altman 法评估两种设备测量值的一致性。
IOLMaster 700 对平均角膜曲率(Km)(P≤0.018)、CCT(P≤0.027)和 ACD(P≤0.001)测量的 ICC 较高,RC 较低,而 AL-Scan 在 2.4mm 区 J45 散光矢量分量的 ICC 较高,RC 较低(P≤0.034)。两种设备在眼轴长度测量方面均具有极好的重复性(ICC=0.999)。两种设备在 Km、CCT、ACD 和 WTW 方面存在系统差异(P≤0.018)。在 2.4mm 区和 3.3mm 区分别测量的 Km 平均差值为-0.196 和-0.144D。相应的 CCT、ACD 和 WTW 距离的平均差值分别为 14.92μm、-0.017mm 和 0.283mm。这些差异导致眼内人工晶状体屈光力预测的统计学显著但临床无显著差异。
本研究表明白内障眼中扫频源生物测量仪和 Scheimpflug 生物测量仪在前节测量重复性和一致性方面存在显著差异。