Comparison of different Raman spectrometer models in the quantification of blood serum analytes.

作者信息

Giana Hector Enrique, Souza Luisa Oliveira, Silveira Landulfo

机构信息

Universidade Anhembi Morumbi, São José dos Campos, Brazil.

Laboratório LabVivalle, São José dos Campos, Brazil.

出版信息

Lasers Med Sci. 2025 Mar 17;40(1):143. doi: 10.1007/s10103-025-04413-y.

Abstract

The analytical performance of three commercial Raman spectrometers in determining the concentrations of analytes associated with diagnosing metabolic, cardiovascular, and renal disorders has been compared. Two portable and one benchtop spectrometers were tested to predict the serum concentrations of triglycerides (TRI), cholesterol (COL), high-density cholesterol, creatinine, urea, and glucose in 193 serum samples using partial least squares (PLS) regression and PLS discriminant analysis, the latter for classifying samples as either altered or within reference values. Strong correlations (r > 0.81) were obtained for TRI and COL analytes using the benchtop and one of the portable spectrometers, and the classification accuracy rates exceeded 90%, suggesting potential for use in clinical screening. Adding a glucose solution improved the correlation and the root mean square error for TRI and COL analytes; however, it did not enhance the correlation or error for the glucose analyte. Raman spectroscopy showed potential to support routine laboratory activities and may have applications in clinical screening.

摘要

比较了三种商用拉曼光谱仪在测定与诊断代谢、心血管和肾脏疾病相关分析物浓度方面的分析性能。测试了两台便携式光谱仪和一台台式光谱仪,使用偏最小二乘法(PLS)回归和PLS判别分析来预测193份血清样本中甘油三酯(TRI)、胆固醇(COL)、高密度胆固醇、肌酐、尿素和葡萄糖的血清浓度,后者用于将样本分类为改变或在参考值范围内。使用台式光谱仪和其中一台便携式光谱仪对TRI和COL分析物获得了强相关性(r>0.81),分类准确率超过90%,表明在临床筛查中有应用潜力。添加葡萄糖溶液改善了TRI和COL分析物的相关性和均方根误差;然而,它并没有提高葡萄糖分析物的相关性或误差。拉曼光谱显示出支持常规实验室活动的潜力,并且可能在临床筛查中有应用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验