一种使用微型计算机断层扫描(Micro-CT)对根管充填牙进行空洞检测的体外实验方案。

A Protocol for Void Detection in Root-filled Teeth Using Micro-CT: Ex-vivo.

作者信息

Gharib Iad, Wong Ferranti S, Davis Graham Roy

机构信息

Department of Restorative Dentistry, University of Dundee, School of Dentistry, Dundee, United Kingdom.

Centre for Oral Bioengineering, Queen Mary University of London, Faculty of Medicine and Dentistry, London, United Kingdom.

出版信息

Eur Endod J. 2025 Jan;10(1):11-17. doi: 10.14744/eej.2023.37167.

Abstract

OBJECTIVE

X-ray microtomography (micro-CT or XMT) has previously been used to measure residual voids in root fillings. However, there is no agreement on a protocol that critically identifies and attempts to solve artefacts inherent to the micro-computed tomography technique. This article aims to describe a protocol for automated detection of voids within root-filled canals taking into account the inherent artefacts, with special interest in the partial volume effect. This is to reduce human errors and increase the accuracy and efficiency of void detection.

METHODS

Human maxillary premolars (n=33) were shaped, cleaned and root-filled using the cold lateral condensation (CLC) technique. Voids were identified using either individual tomographic slices or the new proposed protocol in which: (1) pre-obturation XMT slices were used to identify the coordinates of the canal space; (2) the post-obturation data sets were aligned to the pre-obturation data sets; (3) the voids were identified as voxels with a grey level below a set threshold after subtraction of pre-obturation from post-obturation data sets. A comparison of the voids from these two methods was made.

RESULTS

The visual inspection of slice by slice of the scanned data resulted in full agreement between the tomographic slices and the results gained from the proposed protocol. This confirmed that this protocol provided an automated, effective and accurate method for detecting voids in root-filled canals.

CONCLUSION

The proposed protocol provides an automated method to eliminate inaccuracies from XMT artefacts so that accurate volumetric measurements can be easily obtained. (EEJ-2024-02-031).

摘要

目的

此前已使用X射线显微断层扫描(显微CT或XMT)来测量根管充填物中的残余空隙。然而,对于一种能严格识别并试图解决显微计算机断层扫描技术固有伪影的方案,目前尚无定论。本文旨在描述一种考虑到固有伪影的根管充填根管内空隙自动检测方案,特别关注部分容积效应。这是为了减少人为误差,提高空隙检测的准确性和效率。

方法

使用冷侧向加压(CLC)技术对33颗人类上颌前磨牙进行塑形、清洁和根管充填。使用单个断层切片或新提出的方案识别空隙,其中:(1)充填前的XMT切片用于确定根管空间的坐标;(2)充填后的数据集与充填前的数据集对齐;(3)在从充填后数据集中减去充填前数据集后,将灰度低于设定阈值的体素识别为空隙。对这两种方法得到的空隙进行比较。

结果

对扫描数据逐片进行目视检查,断层切片与所提方案得到的结果完全一致。这证实了该方案为检测根管充填根管内的空隙提供了一种自动、有效且准确的方法。

结论

所提方案提供了一种自动方法,可消除XMT伪影带来的不准确因素,从而能够轻松获得准确的体积测量结果。(EEJ - 2024 - 02 - 031)