Suppr超能文献

范围不确定性对调强混合电子 - 光子放射治疗的影响。

Impact of range uncertainties for intensity-modulated mixed electron-photon radiation therapy.

作者信息

Heng Veng Jean, Renaud Marc-André, Serban Monica, Seuntjens Jan

机构信息

Department of Physics, McGill University, Montreal, Canada.

Medical Physics Unit, McGill University, Montreal, Canada.

出版信息

Med Phys. 2025 Jun;52(6):4785-4792. doi: 10.1002/mp.17771. Epub 2025 Apr 1.

Abstract

BACKGROUND

In the context of mixed electron-photon radiation therapy (MBRT), while the necessity of robust optimization to setup uncertainties is well-established, range uncertainties have yet to be investigated.

PURPOSE

This study provides the first assessment of the impact of range uncertainties on MBRT plans.

METHODS

The percent depth dose of 2 electron beams (6 MeV and 20 MeV) and 1 photon beam (6 MV) are calculated by Monte Carlo using EGSnrc in slab phantoms. Range uncertainties are simulated by generating two copies of the phantom with each voxel's mass density upscaled or downscaled by 3.5%. Two clinical plans for a leg sarcoma case and a post-mastectomy breast case were replanned with MBRT with 2 optimization methods: once without robust optimization and once with robust to both setup and range uncertainties.

RESULTS

Dose discrepancies between the percent depth doses of density-scaled phantoms and the nominal phantom were found to be much larger for electron beams than photons with maximum differences of 6.9% (6 MeV) and 5.5% (20 MeV) versus 1.6% (6 MV) of the maximum dose. In both clinical cases, the region of largest dose discrepancy between the range and nominal scenarios was found to be along the electron's beam path, starting immediately downstream from the target and within a few cm. Even without robust optimization, dose-volume histograms (DVHs) of all relevant structures were not meaningfully degraded under range scenarios. In the breast plan, the ipsilateral lung's V20Gy increased by 1.9% under the worst range scenario. No substantial change in the DVHs of range scenarios were observed between the robustly and non-robustly optimized plan.

CONCLUSION

In the two investigated cases, robustness to range uncertainties was not improved in robustly optimized versus non-robustly optimized plans. A larger study comprising more patients and treatment sites remains to be performed to adequately assess the necessity of robust optimization of MBRT plans to range uncertainties.

摘要

背景

在混合电子 - 光子放射治疗(MBRT)的背景下,虽然针对摆位不确定性进行稳健优化的必要性已得到充分证实,但射程不确定性尚未得到研究。

目的

本研究首次评估了射程不确定性对MBRT计划的影响。

方法

使用EGSnrc通过蒙特卡罗方法在平板体模中计算2个电子束(6 MeV和20 MeV)和1个光子束(6 MV)的百分深度剂量。通过生成两份体模副本模拟射程不确定性,每个体素的质量密度上调或下调3.5%。使用两种优化方法对一例腿部肉瘤病例和一例乳房切除术后乳腺癌病例的两个临床计划进行重新规划:一次不进行稳健优化,一次对摆位和射程不确定性均进行稳健优化。

结果

发现密度缩放体模与标称体模的百分深度剂量之间的剂量差异对于电子束而言远大于光子束,最大差异分别为6.9%(6 MeV)和5.5%(20 MeV),而最大剂量的差异为1.6%(6 MV)。在两个临床病例中,射程与标称情况之间剂量差异最大的区域均位于电子束路径上,从靶区下游紧邻位置开始且在几厘米范围内。即使不进行稳健优化,在射程情况下所有相关结构的剂量体积直方图(DVH)也没有明显恶化。在乳腺癌计划中,在最差射程情况下同侧肺的V20Gy增加了1.9%。在稳健优化和非稳健优化计划之间,未观察到射程情况下DVH有实质性变化。

结论

在这两个研究病例中,表示针对射程不确定性的稳健性方面,稳健优化计划与非稳健优化计划相比并未得到改善。仍需开展一项纳入更多患者和治疗部位的更大规模研究,以充分评估对MBRT计划针对射程不确定性进行稳健优化的必要性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验