Suppr超能文献

胰腺癌电场治疗中电极位置的优化

Optimization of electrode position in electric field treatment for pancreatic cancer.

作者信息

Kim Sangcheol, Ko Yousun, Shin Dongho, Kim Haksoo, Lee Sung Uk, Kim Jonghyun, Kim Tae Hyun, Yoon Myonggeun

机构信息

Department of Biomedical Engineering, Korea University, Seoul, Republic of Korea.

Proton Therapy Center, National Cancer Center, Seoul, Republic of Korea.

出版信息

BMC Gastroenterol. 2025 Apr 4;25(1):222. doi: 10.1186/s12876-025-03807-0.

Abstract

BACKGROUND

In electric field-based cancer treatment, the intensity of the electric field applied to the tumor depends on the position of the electrode array, directly affecting the efficacy of treatment. The present study evaluated the effects of changing the position of the electrode array on the efficacy of electric field treatment for pancreatic cancer.

METHODS

A 3D model was created based on computed tomography images of 13 pancreatic cancer patients. An electrode array was placed on the surface of the model at various positions, and the electric field was calculated for each. Six treatment plans were created for each patient by rotating each electrode array ± 15⁰, ± 30⁰ in the axial plane, and ± 10⁰ in the sagittal plane relative to the reference plan. The frequency was set at 150 kHz and the current density at 31 mArms/cm for calculation of all treatment plans. The mean electric field, minimum electric field, homogeneity index (HI) and coverage index (CI) calculated from the six simulated plans were compared with the reference plan to evaluate the effects of each simulated plan on the tumor.

RESULTS

Comparisons of the simulated plans for each patient with the reference plan showed differences of -2.61 ∼ 11.31% in the mean electric field, -7.03 ∼ 13.87% in the minimum electric field, -64.14 ∼ 13.12% in the HI, and - 24.23 ∼ 11.00% in the CI. Compared with the reference plan, the optimal plans created by changing the electrode position improved the mean electric field 7.41%, the minimum electric field 7.20%, the HI 4.57%, and the CI 8.46%.

CONCLUSIONS

Use of a treatment planning system to determine the optimal placement of the electrode array based on the anatomical characteristics of each patient can improve the intensity of the electric field applied to the tumor.

摘要

背景

在基于电场的癌症治疗中,施加于肿瘤的电场强度取决于电极阵列的位置,直接影响治疗效果。本研究评估了改变电极阵列位置对胰腺癌电场治疗效果的影响。

方法

基于13例胰腺癌患者的计算机断层扫描图像创建三维模型。将电极阵列放置在模型表面的不同位置,并计算每个位置的电场。通过将每个电极阵列相对于参考计划在轴向平面旋转±15⁰、±30⁰,在矢状平面旋转±10⁰,为每位患者创建六个治疗计划。所有治疗计划的计算中,频率设定为150kHz,电流密度设定为31mArms/cm。将六个模拟计划计算得到的平均电场、最小电场、均匀性指数(HI)和覆盖指数(CI)与参考计划进行比较,以评估每个模拟计划对肿瘤的影响。

结果

将每位患者的模拟计划与参考计划进行比较,平均电场的差异为-2.61%至11.31%,最小电场的差异为-7.03%至13.87%,HI的差异为-64.14%至13.12%,CI的差异为-24.23%至11.00%。与参考计划相比,通过改变电极位置创建的最佳计划使平均电场提高了7.41%,最小电场提高了7.20%,HI提高了4.57%,CI提高了8.46%。

结论

使用治疗计划系统根据每位患者的解剖特征确定电极阵列的最佳放置位置,可以提高施加于肿瘤的电场强度。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e378/11969793/5612c47bc773/12876_2025_3807_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验