Shimamoto Yusaku, Morita Yoshinori, Matsunaga Toru, Hino Shiori, Sato Takao, Takao Madoka, Kodama Yuzo
Division of Gastroenterology, Department of Internal Medicine, Kobe University Graduate School of Medicine, Kobe, Japan.
Division of Gastroenterology, International Clinical Cancer Research Center, Kobe University, 1-5-1, Minatojima minami machi, Chuo-ku, Kobe, 650-0047, Japan.
Sci Rep. 2025 Apr 25;15(1):14420. doi: 10.1038/s41598-025-92118-1.
The contamination of the endoscope tip lens results in poor endoscopic imaging. To prevent the adsorption of contaminants, we applied a soft contact lens made of 2-hydroxyethyl methacrylate hydrogel to the tip of a gastrointestinal endoscope. Additionally, a novel endoscopic hood with an upper surface composed of a single layer inside the hood cavity was developed. For in vitro anti-fouling performance testing using blood, changes in light transmittance were examined with hydrogel plates and compared to conventional quartz glass and surfactant-coated glass. The transmittance changes at a wavelength of 415 nm were 1.5% for the hydrogel plate, significantly lower than 82.9% for quartz glass and 85.2% for surfactant-coated glass (p < 0.001), demonstrating the hydrogel's high anti-fouling performance. Another in vitro quantitative assay, evaluated by exposure to blood, lipid, and mucin liquids, showed high anti-fouling performance in the order hydrogel plate > surfactant-coated glass > quartz glass. Moreover, an in vivo pig model performing gastric endoscopic submucosal dissection was used to evaluate anti-fouling performance by analyzing endoscopic images with Michelson contrast. The contrast value of the upper surface group (Hydrogel) was 0.97, significantly higher than 0.62 for the non-upper surface group (Surfactant) (P < 0.001). Therefore, the novel anti-fouling endoscope tip hood could be applied in future clinical applications.
内镜顶端透镜的污染会导致内镜成像不佳。为防止污染物吸附,我们将由甲基丙烯酸2-羟乙酯水凝胶制成的软性隐形眼镜应用于胃肠内镜的顶端。此外,还开发了一种新型内镜罩,其罩腔内表面由单层组成。对于使用血液进行的体外防污性能测试,通过水凝胶板检测透光率的变化,并与传统石英玻璃和表面活性剂涂层玻璃进行比较。水凝胶板在415nm波长处的透光率变化为1.5%,显著低于石英玻璃的82.9%和表面活性剂涂层玻璃的85.2%(p<0.001),表明水凝胶具有很高的防污性能。另一种通过暴露于血液、脂质和粘蛋白液体进行评估的体外定量分析表明,防污性能从高到低依次为水凝胶板>表面活性剂涂层玻璃>石英玻璃。此外,使用进行胃内镜黏膜下剥离的体内猪模型,通过用迈克尔逊对比度分析内镜图像来评估防污性能。上表面组(水凝胶)的对比度值为0.97,显著高于非上表面组(表面活性剂)的0.62(P<0.001)。因此,这种新型防污内镜顶端罩可应用于未来的临床应用。