DeGrella R F, Combs G L, Coffee E E, Stern C H, Frenkel D, Kureshy F, Kanewske W J, Hiatt D A, Shanks R E, Pumphrey J G
Clin Chem. 1985 Sep;31(9):1474-7.
A self-contained light-scattering accessory carousel has been developed for the Abbott TDx fluorescence polarization analyzer, extending the instrument's capabilities to nephelometric methods of analysis, including assays for specific proteins by immunoprecipitation: IgG, IgA, IgM, and transferrin. The scattered light from a green-light-emitting diode (peak wavelength 565 nm) is measured at an angle of 37.5 degrees by the existing optical detection system of the TDx analyzer without modification of the instrument. Measurements are made at quasi-equilibrium, with sample blank correction. No sample pretreatment is required, and antigen-excess is checked automatically. CVs range from 3 to 6% (within-run) and 7 to 9%. (total). Calibration curves may be stored for at least two weeks. A nephelometric method for monitoring chromogenic reactions in the green wavelength region is also described. This method--Scattered Energy Attenuation (SEA)--has been used in preliminary experiments to measure calcium, total protein, iron, and bilirubin.
已为雅培TDx荧光偏振分析仪开发了一种独立的光散射附件转盘,将该仪器的功能扩展到比浊分析方法,包括通过免疫沉淀法检测特定蛋白质:IgG、IgA、IgM和转铁蛋白。来自绿色发光二极管(峰值波长565 nm)的散射光由TDx分析仪现有的光学检测系统在37.5度角处测量,无需对仪器进行修改。在准平衡状态下进行测量,并进行样品空白校正。无需样品预处理,可自动检查抗原过量情况。变异系数在3%至6%(批内)和7%至9%(总)之间。校准曲线可存储至少两周。还描述了一种用于监测绿色波长区域显色反应的比浊法。这种方法——散射能量衰减(SEA)——已用于初步实验中测量钙、总蛋白、铁和胆红素。