Open-source quality assurance for multi-parametric MRI: a diffusion analysis update for the magnetic resonance biomarker assessment software (MR-BIAS).

作者信息

Korte James C, Norris Stanley A, Carr Madeline E, Holloway Lois, Cahoon Glenn D, Neijndorff Ben, van Houdt Petra, Franich Rick

机构信息

Department of Physical Sciences, Peter MacCallum Cancer Centre, 305 Grattan Street, Melbourne, VIC, 3000, Australia.

Department of Biomedical Engineering, The University of Melbourne, Melbourne, Australia.

出版信息

MAGMA. 2025 Apr 26. doi: 10.1007/s10334-025-01252-4.

Abstract

OBJECTIVE

To validate the automated analysis of magnetic resonance imaging (MRI) diffusion phantoms with an updated version of the magnetic resonance biomarker assessment software (MR-BIAS), an open-source tool initially developed for the analysis of MRI relaxometry phantoms.

MATERIALS AND METHODS

The updated MR-BIAS was validated against two published diffusion weighted MRI datasets: (i) a single-site study (n = 48) was used for validation of apparent diffusion coefficients (ADC) and to identify optimal region of interest (ROI) selection, and (ii) a multi-centre multi-vendor study including diffusion imaging from a shared benchmark protocol (n = 49) and site-specific protocols (n = 43). ADC analysis compared both datasets with ROIs manually matched to the original studies, and with automatically detected optimal ROIs.

RESULTS

MR-BIAS ADC values were statistically equivalent (p < 0.05) to original studies within tolerances (manual ROI, automatic ROI) for the single-site study (± 0.01, ± 6 μm/s) and for the multi-vendor study for benchmark (± 4, ± 7 μm/s) and site-specific (± 3, ± 6 μm/s) protocols. The optimal ROI was a central cylinder (height = 10mm, diameter = 10mm). MR-BIAS ADC summary metrics were comparable to those of the original studies.

DISCUSSION

MR-BIAS can automatically and accurately perform ADC analysis of diffusion phantoms, making the software suitable for the quality assurance of multi-centre studies of multi-parametric MRI.

摘要

目的

使用磁共振生物标志物评估软件(MR-BIAS)的更新版本验证磁共振成像(MRI)扩散体模的自动分析,MR-BIAS是最初为分析MRI弛豫测量体模而开发的开源工具。

材料与方法

针对两个已发表的扩散加权MRI数据集对更新后的MR-BIAS进行验证:(i)一项单中心研究(n = 48)用于表观扩散系数(ADC)的验证和确定最佳感兴趣区域(ROI)选择,以及(ii)一项多中心多供应商研究,包括来自共享基准协议(n = 49)和特定地点协议(n = 43)的扩散成像。ADC分析将两个数据集与手动匹配原始研究的ROI以及自动检测到的最佳ROI进行比较。

结果

对于单中心研究(±0.01,±6μm/s)以及多供应商研究的基准协议(±4,±7μm/s)和特定地点协议(±3,±6μm/s),MR-BIAS的ADC值在公差范围内(手动ROI、自动ROI)与原始研究在统计学上等效(p < 0.05)。最佳ROI是一个中心圆柱体(高度 = 10mm,直径 = 10mm)。MR-BIAS的ADC汇总指标与原始研究的指标相当。

讨论

MR-BIAS可以自动且准确地对扩散体模进行ADC分析,使该软件适用于多参数MRI多中心研究的质量保证。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验