使用锥束CT和显微CT图像评估小梁骨微观结构和皮质形态:管电压设置的影响

Evaluation of trabecular bone microstructure and cortical morphology using cone-beam and micro-CT images: impact of tube voltage setting.

作者信息

Baydar Oğuzhan, Onem Erinc, Mert Ali, Baksi B Guniz

机构信息

Faculty of Dentistry, Department of Oral and Maxillofacial Radiology, Eskisehir Osmangazi University, Eskişehir, Turkey.

School of Dentistry, Department of Oral and Maxillofacial Radiology, Ege University, Izmir, Turkey.

出版信息

BMC Oral Health. 2025 Apr 29;25(1):664. doi: 10.1186/s12903-025-05855-0.

Abstract

OBJECTIVE

To evaluate the effect of various tube voltage (kV) settings on the accuracy of cone-beam computed tomography (CBCT) images in measuring trabecular microstructure and cortical morphology, using micro-CT (μCT) as the reference.

METHODS

Ten bone samples of sheep mandibles were scanned using both μCT and CBCT at three different tube voltage settings (80, 85, and 90 kV). Identical regions of interest (ROIs) on trabecular and cortical bones were analyzed in all images. Measurements of trabecular microstructure included bone volume fraction (BV/TV), trabecular thickness (Tb.Th) and space (Tb.Sp), while thickness (Ct.Th) and area (Ct.Ar) of cortical bone were measured to determine cortical morphology. Measurements were compared using paired t-test, while agreement between measurements of two modalities was assessed using Bland-Altman analysis. One-way ANOVA was used to determine differences in measurements of CBCT images at different kVs (p < 0.01).

RESULTS

Compared to μCT, CBCT overestimated trabecular parameters and Ct.Th but underestimated Ct.Ar, with high agreement observed between the methods. Significant differences were found for all measurements except BV/TV and Ct.Ar at all kVs. No differences were observed between CBCT measurements at different tube potentials.

CONCLUSION

The tube voltage of CBCT has minimal impact on the measurement accuracy of most microstructural parameters. BV/TV and Ct.Ar measurements may be particularly preferred for bone evaluations using CBCT images.

摘要

目的

以微型计算机断层扫描(μCT)为参照,评估不同管电压(kV)设置对锥形束计算机断层扫描(CBCT)图像测量小梁微结构和皮质形态准确性的影响。

方法

使用μCT和CBCT在三种不同管电压设置(80、85和90 kV)下对10个绵羊下颌骨样本进行扫描。在所有图像中分析小梁骨和皮质骨上相同的感兴趣区域(ROI)。小梁微结构的测量包括骨体积分数(BV/TV)、小梁厚度(Tb.Th)和间距(Tb.Sp),同时测量皮质骨的厚度(Ct.Th)和面积(Ct.Ar)以确定皮质形态。测量结果采用配对t检验进行比较,而两种成像方式测量结果之间的一致性则使用布兰德-奥特曼分析进行评估。采用单因素方差分析确定不同kV下CBCT图像测量结果的差异(p < 0.01)。

结果

与μCT相比,CBCT高估了小梁参数和Ct.Th,但低估了Ct.Ar,两种方法之间观察到高度一致性。在所有kV下,除BV/TV和Ct.Ar外,所有测量结果均存在显著差异。不同管电压下CBCT测量结果之间未观察到差异。

结论

CBCT的管电压对大多数微结构参数的测量准确性影响极小。使用CBCT图像进行骨评估时,BV/TV和Ct.Ar测量可能特别适用。