Reduction of artifacts from contrast media in spectral-detector CT by combined use of virtual monoenergetic images and orthopedic metal-artifact-reduction algorithm.

作者信息

Pan Nannan, Liu Junpeng, Wang Shuo, Miao Zhenwei

机构信息

Department of Radiology, Tianjin Medical University Baodi Hospital, Tianjin, China.

Department of Orthopedic Surgery, Tianjin Medical University Baodi Hospital, Tianjin, China.

出版信息

Medicine (Baltimore). 2025 May 16;104(20):e42370. doi: 10.1097/MD.0000000000042370.

Abstract

This study assesses the artifacts reduction caused by contrast media (CM) in the subclavian and axillary veins in dual-layer spectral-detector CT using virtual monoenergetic images (VMI) and orthopedic metal-artifact-reduction (O-MAR) algorithm. A total of 61 nonconsecutive patients who underwent enhanced chest CT examinations were enrolled in the present study. Conventional images (CI), VMI, CI with O-MAR algorithm (CI + O-MAR), and VMI with O-MAR algorithm (VMI + O-MAR) were reconstructed using the same arterial CT dataset. The VMI and VMI + O-MAR images were reconstructed at 130 keV. Subjective image quality assessment was performed in terms of degree of artifacts and overall image quality using the Likert-scale. The differences in image noise, artifact index (AI) and CT number difference were compared among the 4 reconstructed images. Soft tissue adjacent to CM exhibited fewer artifacts and better image quality in VMI + O-MAR images than in VMI, CI + O-MAR images, and CI (P < .001). In addition, image noise and AI in VMI + O-MAR images were lower than those in VMI and CI + O-MAR images (P < .001). CT number difference was lower in VMI and VMI + O-MAR images than in CI + O-MAR images (P < .01). The CI had the highest values in image noise, AI and CT number difference (P < .001). The combination of O-MAR and 130 keV VMI showed a significant reduction of artifacts from CM than each technique alone and provided better image quality and diagnostic confidence. The combined application of O-MAR and 130 keV VMI can be a better alternative to reduce artifacts from CM in spectral-detector CT.

摘要

本研究评估了在双层光谱探测器CT中,使用虚拟单能图像(VMI)和骨科金属伪影减少(O-MAR)算法时,对比剂(CM)在锁骨下静脉和腋静脉中所导致的伪影减少情况。本研究共纳入61例接受增强胸部CT检查的非连续患者。使用相同的动脉CT数据集重建常规图像(CI)、VMI、采用O-MAR算法的CI(CI + O-MAR)以及采用O-MAR算法的VMI(VMI + O-MAR)。VMI和VMI + O-MAR图像在130 keV下重建。使用李克特量表,从伪影程度和整体图像质量方面进行主观图像质量评估。比较了4种重建图像在图像噪声、伪影指数(AI)和CT值差异方面的不同。与VMI、CI + O-MAR图像和CI相比,CM相邻的软组织在VMI + O-MAR图像中表现出更少的伪影和更好的图像质量(P <.001)。此外,VMI + O-MAR图像中的图像噪声和AI低于VMI和CI + O-MAR图像(P <.001)。VMI和VMI + O-MAR图像中的CT值差异低于CI + O-MAR图像(P <.01)。CI在图像噪声、AI和CT值差异方面具有最高值(P <.001)。O-MAR与130 keV VMI的组合比单独使用每种技术能更显著地减少CM产生的伪影,并提供更好的图像质量和诊断信心。O-MAR与130 keV VMI的联合应用可能是减少光谱探测器CT中CM伪影的更好选择。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fcfb/12091650/e7fe0d990886/medi-104-e42370-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验