用于估计CT中病变可检测性的患者特异性通道化霍特林观察者。

Patient-specific Channelized Hotelling observer to estimate lesion detectability in CT.

作者信息

Zhou Zhongxing, Wellinghoff Jarod, McCollough Cynthia H, Yu Lifeng

机构信息

Department of Radiology, Mayo Clinic, Rochester, MN, 55905, USA.

出版信息

Proc SPIE Int Soc Opt Eng. 2025 Feb;13405. doi: 10.1117/12.3047381. Epub 2025 Apr 8.

Abstract

Task-based image quality assessment is essential for CT protocol and radiation dose optimization. Despite many ongoing efforts, there is still an unmet need to measure and monitor the quality of images acquired from each patient exam. In this work, we developed a patient-specific channelized Hotelling observer (CHO)-based method to estimate the lesion detectability for each individual patient scan. The ensemble of background was created from patient images to include both relatively uniform regions and anatomically varying regions. Signals were modelled from lesions of different sizes and contrast levels after incorporating the effect of contrast-dependent spatial resolution. Index of detectability (d') was estimated using a CHO framework. This method was applied to clinical patient images obtained from a CT scanner at 3 different radiation dose levels. The d' for 5 different lesion size/contrast conditions was calculated across the scan range of each patient exam. The average noise levels and the d' averaged from 5 conditions were 13.2/3.78, 17.1/2.93 and 21.9/2.43 at 100%, 50% and 25% dose levels, respectively.

摘要

基于任务的图像质量评估对于CT协议和辐射剂量优化至关重要。尽管正在进行许多努力,但仍存在未满足的需求,即测量和监测从每位患者检查中获取的图像质量。在这项工作中,我们开发了一种基于患者特定的通道化霍特林观察者(CHO)的方法,以估计每位患者扫描的病变可检测性。背景集合由患者图像创建,包括相对均匀的区域和解剖结构变化的区域。在纳入对比度依赖空间分辨率的影响后,对不同大小和对比度水平的病变进行信号建模。使用CHO框架估计可检测性指数(d')。该方法应用于从CT扫描仪在3种不同辐射剂量水平下获得的临床患者图像。在每位患者检查的扫描范围内,计算了5种不同病变大小/对比度条件下的d'。在100%、50%和25%剂量水平下,5种条件下的平均噪声水平和d'平均值分别为13.2/3.78、17.1/2.93和21.9/2.43。

相似文献

1
Patient-specific Channelized Hotelling observer to estimate lesion detectability in CT.
Proc SPIE Int Soc Opt Eng. 2025 Feb;13405. doi: 10.1117/12.3047381. Epub 2025 Apr 8.
7
Use of a channelized Hotelling observer to assess CT image quality and optimize dose reduction for iteratively reconstructed images.
J Med Imaging (Bellingham). 2017 Jul;4(3):031213. doi: 10.1117/1.JMI.4.3.031213. Epub 2017 Oct 3.
10
Low-contrast detectability of photon-counting-detector CT at different scan modes and image types in comparison with energy-integrating-detector CT.
J Med Imaging (Bellingham). 2024 Dec;11(Suppl 1):S12803. doi: 10.1117/1.JMI.11.S1.S12803. Epub 2024 May 24.

本文引用的文献

1
Development and validation of a noise insertion algorithm for photon-counting-detector CT.
Med Phys. 2024 Sep;51(9):5943-5953. doi: 10.1002/mp.17263. Epub 2024 Jun 23.
5
6
Development and validation of an automated methodology to assess perceptual noise texture in liver CT.
J Med Imaging (Bellingham). 2021 Sep;8(5):052113. doi: 10.1117/1.JMI.8.5.052113. Epub 2021 Oct 25.
7
A Web-Based Software Platform for Efficient and Quantitative CT Image Quality Assessment and Protocol Optimization.
Proc SPIE Int Soc Opt Eng. 2021 Feb;11595. doi: 10.1117/12.2582123. Epub 2021 Feb 15.
8
Observer Performance with Varying Radiation Dose and Reconstruction Methods for Detection of Hepatic Metastases.
Radiology. 2018 Nov;289(2):455-464. doi: 10.1148/radiol.2018180125. Epub 2018 Sep 11.
9
Estimating detectability index : development and validation of an automated methodology.
J Med Imaging (Bellingham). 2018 Jul;5(3):031403. doi: 10.1117/1.JMI.5.3.031403. Epub 2017 Dec 11.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验