评估步行运动期间用于测量心脏代谢指标的PNOE¯的有效性和可靠性。

Assessing the Validity and Reliability of the PNOE¯ for Measuring Cardiometabolic Outcomes During Walking Exercise.

作者信息

Kwok Manny M Y, Ng Shamay S M, Myers Jonathan, So Billy C L

机构信息

Department of Rehabilitation Sciences, The Hong Kong Polytechnic University, Hong Kong, China.

VA Palo Alto Health Care System, Stanford University, Palo Alto, CA 94304, USA.

出版信息

J Funct Morphol Kinesiol. 2025 May 6;10(2):159. doi: 10.3390/jfmk10020159.

Abstract

The accuracy of measurement of cardiometabolic outcomes in terms of gaseous exchange and energy expenditure of individuals is crucial. The objective of this study was to compare the validity and reliability of the PNOE¯ in measuring cardiometabolic outcomes from the respiratory gaseous exchange of healthy individuals during treadmill walking exercise. A total of 21 healthy subjects (15 male and 6 female) aged 22.76 ± 3.85 years took part in this study. Oxygen uptake (VO), carbon dioxide production (VCO), respiratory exchange ratio (RER), metabolic equivalents (METs), tidal volume (VT), and energy expenditure (EE) were measured using the PNOE¯ and COSMED K5 portable systems during a twenty-eight-minute, four-stage incremental protocol, where speed increased from 1.7 mph to 4.2 mph with a 2% incline on a treadmill. Test-retest reliability was tested on separate days with trail repetition. Validity was evaluated by Bland-Altman plots, intraclass correlation coefficients (ICCs) and mean percentage difference. ICCs showed that VCO was in the good range (0.75-0.90). The ICC of the RER from stages 1 to 3 of the incremental protocol and the VT from stages 2 to 4 of the incremental protocol showed good to excellent reliability. No clear trend was seen for VO, VCO, and EE datapoints with variations in speed. Pearson's correlation coefficients were moderately high (r = 0.60-0.79) between VO, VCO, RER, METs, VT, and EE measured by the PNOE¯ and K5 systems. All subjects, except for a few cases in VT, were within the upper and lower 95% confidence intervals of the acceptable range of the Bland-Altman plots. The PNOE¯ system is a valid and reliable measure of cardiometabolic outcomes and is comparable to the COSMED K5 system.

摘要

就个体的气体交换和能量消耗而言,测量心脏代谢结果的准确性至关重要。本研究的目的是比较PNOE¯在测量健康个体在跑步机步行运动期间呼吸气体交换的心脏代谢结果方面的有效性和可靠性。共有21名年龄在22.76±3.85岁的健康受试者(15名男性和6名女性)参与了本研究。在一个28分钟的四阶段递增方案中,使用PNOE¯和COSMED K5便携式系统测量了摄氧量(VO)、二氧化碳产生量(VCO)、呼吸交换率(RER)、代谢当量(METs)、潮气量(VT)和能量消耗(EE),其中跑步机速度从1.7英里/小时增加到4.2英里/小时,坡度为2%。在不同日期进行重复测试以检验重测信度。通过Bland-Altman图、组内相关系数(ICC)和平均百分比差异来评估效度。ICC显示VCO处于良好范围(0.75 - 0.90)。递增方案第1至3阶段的RER以及递增方案第2至4阶段的VT的ICC显示出良好至优秀的信度。VO、VCO和EE数据点未随速度变化呈现明显趋势。PNOE¯和K5系统测量的VO、VCO、RER、METs、VT和EE之间的皮尔逊相关系数中等偏高(r = 0.60 - 0.79)。除VT中的少数情况外,所有受试者均在Bland-Altman图可接受范围的上下95%置信区间内。PNOE¯系统是一种有效且可靠的心脏代谢结果测量方法,与COSMED K5系统相当。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5fd4/12101258/b5bf69a05c99/jfmk-10-00159-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验