Hong Seojin, Choi Hyun, Kweon Hyosun
Department of Clinical Rehabilitation Research, Rehabilitation Research Institute, Korea National Rehabilitation Center, 58 Samgaksan-ro, Gangbuk-gu, Seoul, 01022, Republic of Korea, 82 02-901-1966.
Department of Healthcare and Public Health, Rehabilitation Research Institute, Korea National Rehabilitation Center, Seoul, Republic of Korea.
JMIR Hum Factors. 2025 May 30;12:e60607. doi: 10.2196/60607.
Musculoskeletal disorders significantly impair physical function and quality of life, necessitating systematic rehabilitation. Electrically powered orthopedic exercisers, such as continuous passive motion devices, are widely used to enhance joint mobility and muscle recovery. However, existing devices often lack advanced functionalities and user-specific adaptability, limiting their effectiveness. To address these shortcomings, the Rebless Pro was developed as a novel device supporting active and passive exercises with personalized treatment programs.
This study aimed to conduct a formative usability evaluation of the Rebless Pro prototype using focus group interviews (FGIs) and satisfaction surveys with health care professionals specializing in rehabilitation medicine. The goal was to identify areas for improvement to enhance the safety, usability, and information clarity of the device.
Usability evaluation was performed at the National Rehabilitation Center with 10 participants (5 physiatrists and 5 physical therapists) who had prior experience using similar devices. FGIs were conducted to collect qualitative insights into user experiences, while satisfaction surveys provided quantitative data on ease of use of the user interface and identifiability and understanding of information. Data collection focused on identifying risk factors and usability challenges.
Three key areas for improvement were identified: (1) product upgrades to ensure patient safety, including adjustments to exercise speed and resistance; (2) hardware and software improvements to improve usability, including adjustments to the location of the emergency button and improvements to the graphical user interface elements; and (3) improvements to the user manual, including detailed contraindications, patient criteria, and clearer operating instructions. Although the mean score of physiatrists (mean 4.463, SD 0.298) was higher than that of physical therapists (mean 4.114, SD 0.829) in terms of the ease of use of the user interface, the difference was not statistically significant (P=.69). Similarly, in the category of identifiability and understanding of information, higher scores were again reported by physiatrists (mean score 4.467, SD 0.506) than by physical therapists (mean 3.733, SD 0.894), but this difference was also not statistically significant (P=.22).
Usability evaluation provided actionable insights into improving the Rebless Pro's safety, usability, and information clarity. To further refine the device, iterative usability evaluations involving both health care professionals and patients are recommended. These efforts are expected to contribute to the development of a safe, effective, and user-friendly electrically powered orthopedic exerciser suitable for commercialization.
肌肉骨骼疾病会严重损害身体功能和生活质量,因此需要系统的康复治疗。电动骨科锻炼设备,如持续被动运动装置,被广泛用于增强关节活动度和肌肉恢复。然而,现有设备往往缺乏先进功能和针对用户的适应性,限制了其有效性。为解决这些不足,研发了Rebless Pro作为一种支持主动和被动锻炼并具有个性化治疗方案的新型设备。
本研究旨在通过焦点小组访谈(FGI)以及对康复医学专业医护人员进行满意度调查,对Rebless Pro原型进行形成性可用性评估。目标是确定改进领域,以提高设备的安全性、可用性和信息清晰度。
在国家康复中心对10名有使用类似设备经验的参与者(5名物理治疗师和5名物理治疗师)进行了可用性评估。进行焦点小组访谈以收集关于用户体验的定性见解,而满意度调查则提供了关于用户界面易用性以及信息可识别性和理解程度的定量数据。数据收集集中于识别风险因素和可用性挑战。
确定了三个关键改进领域:(1)产品升级以确保患者安全,包括调整锻炼速度和阻力;(2)硬件和软件改进以提高可用性,包括调整紧急按钮位置和改进图形用户界面元素;(3)改进用户手册,包括详细的禁忌证、患者标准和更清晰的操作说明。尽管在用户界面易用性方面,物理治疗师的平均得分(平均4.463,标准差0.298)高于物理治疗师(平均4.11,4,标准差0.829),但差异无统计学意义(P = 0.69)。同样,在信息可识别性和理解类别中,物理治疗师报告的得分(平均得分4.467,标准差0.506)再次高于物理治疗师(平均3.733,标准差0.894),但这种差异也无统计学意义(P = 0.22)。
可用性评估为改进Rebless Pro的安全性、可用性和信息清晰度提供了可操作的见解。为进一步优化该设备,建议对医护人员和患者进行迭代可用性评估。这些努力有望有助于开发出一种适合商业化的安全、有效且用户友好的电动骨科锻炼设备。