Taysi Nuri Mert, Erten Taysi Aysegul, Erçal Pınar, Sismanoglu Soner
İstanbul University-Cerrahpasa, Faculty of Dentistry, Department of Oral and Maxillofacial Surgery, İstanbulTürkiye.
Altinbas University, Faculty of Dentistry, Department of Oral and Maxillofacial Surgery, İstanbulTürkiye.
Eur Oral Res. 2025 Jan 5;59(1):46-52. doi: 10.26650/eor.20241296069.
The objective of this study was to assess the effects of various implant insertion techniques on the primary stability of dental implants in both type II and type IV cadaveric bovine.
A total of 48 dental implants (BEGO Semados RSX, BEGO Implant Systems GmbH and Co. KG, Germany) with a diameter of 3.75 mm and a length of 12 mm were used in the experiments. Bovine bone ribs were adjusted to mimic type II and type IV bone characteristics. Following the preparation of recipient sites, implants were inserted using three different protocols: machine-driven insertion (Standard group, Std group), ratchet insertion (Ratcheted, R Group), and a combination of both (Std + R group). The Osstell® Beacon device was used to record the implant stability quotient (ISQ) of each implant immediately after insertion. Two-way analysis of variance and Bonferroni tests were used for statistical evaluation.
Bone type significantly influenced the ISQ values (p<0.05). However, when comparing insertion protocols separately for type II and type IV bone, no significant differences were observed. In type IV bone, both the Std group and R group exhibited significantly lower ISQ values compared to the same groups in type II bone (p<0.05 for each). Nevertheless, there were no significant differences in the ISQ values when employing the Std+R technique between the two types of bone.
Combining machine-driven and ratchet insertion techniques may prove beneficial in optimizing ISQ values in bovine samples simulating type IV bone.
本研究的目的是评估各种种植体植入技术对II型和IV型尸体牛牙种植体初始稳定性的影响。
实验共使用48颗直径3.75mm、长度12mm的牙种植体(德国贝格公司的BEGO Semados RSX)。对牛肋骨进行调整以模拟II型和IV型骨特征。在制备种植窝后,采用三种不同方案植入种植体:机械驱动植入(标准组,Std组)、棘轮式植入(棘轮组,R组)以及两者结合(Std + R组)。植入后立即使用Osstell® Beacon设备记录每个种植体的种植体稳定性商数(ISQ)。采用双向方差分析和Bonferroni检验进行统计学评估。
骨类型对ISQ值有显著影响(p<0.05)。然而,分别比较II型和IV型骨的植入方案时,未观察到显著差异。在IV型骨中,Std组和R组的ISQ值均显著低于II型骨中相同组的值(每组p<0.05)。然而,在两种骨类型中采用Std+R技术时,ISQ值没有显著差异。
在模拟IV型骨的牛样本中,结合机械驱动和棘轮式植入技术可能有助于优化ISQ值。